Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.
Published June 6, 2018 · Recall initiated: December 20, 2017
Summary
FDA records medical device Class II recall Z-1938-2018, published June 6, 2018 and initiated December 20, 2017. Recalling firm: Ad-Tech Medical Instrument Corporation; Status: Terminated; Reported quantity: 30 devices; Lot: Catalog No. DDK2-2.4-30X, Lot No. 0111664 208140649, UDI No. 01008418231011741719080110011174 5208140649; Catalog No. DDK2-2.8-30X, Lot No. 0111745 208140649, UDI No. 01008418231011671719080110011166 4208140649. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ad-Tech Medical Instrument Corporation
- Distributor
- Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.
- Lot
- Catalog No. DDK2-2.4-30X, Lot No. 0111664 208140649, UDI No. 01008418231011741719080110011174 5208140649; Catalog No. DDK2-2.8-30X, Lot No. 0111745 208140649, UDI No. 01008418231011671719080110011166 4208140649
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth ElectrodeOctober 30, 2024
- FDAAD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth ElectrodeOctober 30, 2024
- FDAAD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth ElectrodeOctober 30, 2024
- FDAAD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4NSeptember 25, 2019
- FDAMEDLINE Sterile* Needle Counter, 40Ct. Foam Block / Adhesive, reorder: DYNJNC40A Product Usage: A needleJune 6, 2018
- FDASymmetry Olsen bipolar cord: (a) Bipolar Cord, Twin Pin Connector to Shrouded Instrument Connector, 12'June 6, 2018
Questions about this recall
Was Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame. recalled?
Yes. FDA published recall Z-1938-2018 on June 6, 2018. Reported hazard: There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.
What should I do if I have this product?
Voluntary: Firm initiated · Letter