FDAFDA · Food and drugs·Recall number Z-1940-2014
Stirling Ultracold Portable Shuttle, ULT-25. Storage for cold-chain management, short-term or long-term ultra-low and storage and transport of biological specimens.
Published July 9, 2014 · Recall initiated: March 20, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The external power supply input is overheating at the connection to the freezer.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Global Cooling Inc
- Distributor
- Worldwide Distribution -- USA, including the states of AK, AR, AZ, CA, CO, CT, FL, GA, GU, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, and WA; and, the countries of Australia, Brazil, Canada, China, India, Indonesia, Japan, Korea, Kuwait, Mexico, Netherlands, Pakistan, Singapore, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.
- Lot
- Codes: ULT-25-255 & ULT-25-713
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.