FDAFDA · Food and drugs·Recall number Z-1941-2014
illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.
Published July 9, 2014 · Recall initiated: April 22, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.
What to do
Voluntary: Firm initiated · Telephone
Recall details
- Manufacturer
- Meridian Bioscience Inc
- Distributor
- Worldwide Distribution - USA including AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, and WV. Internationally to Australia.
- Lot
- Serial Numbers: IP2669, IP2670, IP2671, IP2672, IP2673, IP2674, IP2675, IP2676, IP2677, IP2678, IP2680, IP2682, IP2683, IP2684, IP2685, IP2686, IP2687, IP2688, IP2689, IP2690, IP2691, IP2692, IP2693, IP2694, IP2695, IP2696, IP2697, IP2698, IP2699, IP2700, IP2701, IP2702, IP2703, IP2704, IP2705, IP2706.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States