3.2mm Drill Bit. This is a component within the PediLoc Locking Plate System utilized to pre-drill a hole in bone prior to placement of a bone screw.
Published August 21, 2013 · Recall initiated: July 24, 2013
Summary
FDA records medical device Class II recall Z-1941-2013, published August 21, 2013 and initiated July 24, 2013. Recalling firm: OrthoPediatrics Corp; Status: Terminated; Reported quantity: 68; Lot: LOT 91300335 - Catalogue Number 01-1050-0002. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
OrthoPediatrics has initiated a voluntary recall of Part # 01-1050-002, 2.5mm Drill Bit, Lot #91300335. Only drill bits that are in their original packaging are being recalled. The description on the package of this drill bit has a wrong size in the description (3.2mm). The recall affects only packaged devices still in original packaging due to misbranding of the product. The drill bit itself has
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- OrthoPediatrics Corp
- Distributor
- Nationwide distribution, including the states of MO, MI, WA, FL, NY, TN, CO, AZ, VA, TX, CA, MN, LA, NC, and UT.
- Lot
- LOT 91300335 - Catalogue Number 01-1050-0002
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOrthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended forAugust 5, 2020
- FDAOrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - ProductJuly 15, 2020
- FDAModel No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. ProductNovember 6, 2019
- FDAResponse BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000) is utilized to allow theJune 14, 2017
- FDAGE LightSpeed CT750 HD The GE LightSpeed CT750 HD Computed Tomography X-ray system is intended to produceAugust 21, 2013
- FDAGE Healthcare, Optima XR220amx, Optima XR200amx, and Brivo XR285amx. The series are indicated for use on adultAugust 21, 2013
Questions about this recall
Was 3.2mm Drill Bit. This is a component within the PediLoc Locking Plate System utilized to pre-drill a hole in bone prior to placement of a bone screw. recalled?
Yes. FDA published recall Z-1941-2013 on August 21, 2013. Reported hazard: OrthoPediatrics has initiated a voluntary recall of Part # 01-1050-002, 2.5mm Drill Bit, Lot #91300335. Only drill bits that are in their original packaging are being recalled. The description on the package of this drill bit has a wrong size in the description (3.2mm). The recall affects only packaged devices still in original packaging due to misbranding of the product. The drill bit itself has
What should I do if I have this product?
Voluntary: Firm initiated · Telephone