FDAFDA · Food and drugs·Recall number Z-1944-2012

MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086

Published July 18, 2012 · Recall initiated: June 8, 2012

Summary

FDA records medical device Class II recall Z-1944-2012, published July 18, 2012 and initiated June 8, 2012. Recalling firm: Impac Medical Systems Inc; Status: Terminated; Reported quantity: 11 affected sites; Lot: Version 2.10 - all builds and version 2.20 - builds up to and including 2.20.07A7. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Impac Medical Systems Inc
Distributor
Worldwide Distribution -- USA, including the states of VA, CT, GA, NJ and NY and the countries of Belgium, Denmark, France, and the United Kingdom.
Lot
Version 2.10 - all builds and version 2.20 - builds up to and including 2.20.07A7.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MOSAIQ Oncology Information Systems interfaced to Elekta Treatment machines. An image enabled electronic medical record system used for oncology workflow management. Manufactured by: Impac Medical Systems 100 Mathilda Place, 5th floor Sunnyvale, Ca 94086 recalled?

Yes. FDA published recall Z-1944-2012 on July 18, 2012. Reported hazard: Impac Medical became aware of a problem with the recording of the Wedge Motor Units in MOSAIQ when it resumes after a machine terminates/faults, the remaining Wedge MU was being erroneously set to zero.

What should I do if I have this product?

Voluntary: Firm initiated · Letter