Back to search
FDAFDA · Food and drugs·Recall number Z-1944-2020

iChem¿VELOCITY Urine Chemistry System-

Published May 20, 2020 · Recall initiated: March 30, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
US Distribution: Nationwide Foreign Distribution: Worldwide
Lot
Catalog #700-7176-001 and #700-7177-001, All part numbers
Risk classification
Class II
Area (FDA)
device
Country
United States