FDAFDA · Food and drugs·Recall number Z-1949-2026

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Published May 6, 2026 · Recall initiated: March 23, 2026

Summary

FDA records medical device Class II recall Z-1949-2026, published May 6, 2026 and initiated March 23, 2026. Recalling firm: Hologic, Inc; Status: Ongoing; Reported quantity: 1200 units; Lot: Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hologic, Inc
Distributor
Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
Lot
Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035. recalled?

Yes. FDA published recall Z-1949-2026 on May 6, 2026. Reported hazard: It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

What should I do if I have this product?

Voluntary: Firm initiated · Letter