Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.
Published May 6, 2026 · Recall initiated: March 23, 2026
Summary
FDA records medical device Class II recall Z-1949-2026, published May 6, 2026 and initiated March 23, 2026. Recalling firm: Hologic, Inc; Status: Ongoing; Reported quantity: 1200 units; Lot: Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hologic, Inc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.
- Lot
- Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS,May 20, 2026
- FDASELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D,May 20, 2026
- FDAMedtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatmentMay 6, 2026
- FDABrand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number:May 6, 2026
- FDAHologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guideDecember 18, 2024
- FDABioZorb Marker. Radiographic soft tissue marker.December 4, 2024
Questions about this recall
Was Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035. recalled?
Yes. FDA published recall Z-1949-2026 on May 6, 2026. Reported hazard: It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.
What should I do if I have this product?
Voluntary: Firm initiated · Letter