Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.
Published May 6, 2026 · Recall initiated: March 11, 2026
Summary
FDA records medical device Class II recall Z-1948-2026, published May 6, 2026 and initiated March 11, 2026. Recalling firm: Medtronic, Inc.; Status: Ongoing; Reported quantity: 120698 units; Lot: US: Basic UDI Number 0763000B000262284, GTIN: 00763000871741; 00763000973391; 00763000973506; 00763000993269; 00199150003709. OUS: 00763000871734; 00763000973384; 00763000973513; 00763000811341; 00763000911515; 00763000963675; 00199150003716. Serial Numbers: 227076469, 2301234000, 2301321000, 2301373000, 2301543000, 2301547000, 2301551000, 2301552000, 2301556000, 2301558000, 2301559000, 2301560000, 2301709000, 0012573781, 0012576276, 0012578858, 0012583026, 0012583027, 0012590913, 0012590914, 0012594540, 0012594541, 0012609253, 0012609255, 0012609258, 0012614806, 0012614807, 0012614808, 0012619893, 0012619896, 0012619898, 0012619899, 0012623987, 0012623988, 0012623991, 0012631219, 0012631220, 0012631221, 0012631222, 0012635279, 0012635281, 0012635282, 0012635283, 0012635284, 0012641293, 0012641294, 0012641295, 0012641297, 0012643392, 0012643393, 0012643396, 0012643399, 0012645205, 0012645206, 0012645208, 0012655005, 0012655006, 0012655008, 0012655009, 0012655010, 0012655011, 0012655012, 0012655013, 0012655014, 0012655015, 0012655016, 0012655017, 0012665669, 0012665670, 0012665671, 0012665673, 0012665674, 0012665675, 0012670372, 0012670375, 0012670377, 0012670380, 0012670381, 0012670383, 0012672152, 0012672156, 0012672157, 0012672158, 0012672159, 0012672160, 0012672161, 0012672162, 0012672163, 0012672164, 0012672166, 0012672167, 0012681148, 0012681149, 0012681151, 0012681152, 0012681153, 0012681154, 0012681155, 0012681157, 0012681159, 0012681160, 0012685946, 0012685947, 0012685948, 0012685950, 0012685951, 0012685952, 0012685953, 0012685954, 0012685955, 0012688084, 0012690714, 0012690716, 0012690717, 0012690719, 0012690731, 0012690732, 0012690733, 0012690735, 0012690736, 0012690737, 0012690738, 0012690739, 0012699686, 0012699687, 0012699688, 0012699689, 0012699691, 0012699693, 0012699694, 0012699695, 0012699696, 0012699697, 0012699698, 0012699699, 0012703730, 0012703731, 0012703732, 0012703733, 0012703734, 0012703735, 0012703736, 0012703737, 0012703740, 0012703741, 0012703743, 0012703744, 0012705502, 0012705503, 0012705504, 0012705505, 0012705506, 0012705507, 0012705508, 0012705509, 0012705510, 0012705512, 0012716663, 0012716664, 0012716665, 0012716666, 0012716667, 0012716668, 0012716669, 0012716670, 0012716671, 0012716672, 0012716673, 0012716674, 0012724581, 0012724583, 0012724605, 0012724606, 0012724607, 0012724608, 0012724609, 0012724610, 0012724612, 0012724613, 0012728680, 0012728681, 0012728682, 0012728684, 0012728686, 0012728688, 0012728689, 0012740819, 0012740820, 0012740821, 0012740823, 0012740824, 0012740825, 0012740855, 0012740856, 0012740857, 0012740858, 0012743381, 0012743382, 0012743384, 0012743385, 0012743386, 0012743387, 0012743388, 0012743389, 0012743390, 0012743393, 0012743495, 0012743496, 0012752332, 0012752333, 0012752334, 0012752335, 0012752336, 0012752337, 0012752338, 0012752340, 0012752341, 0012754564, 0012754565, 0012754566, 0012754567, 0012754568, 0012754569, 0012754570, 0012754571, 0012758280, 0012758281, 0012758282, 0012758285, 0012758286, 0012758287, 0012758288, 0012758289, 0012760403, 0012760404, 0012760405, 0012760406, 0012760407, 0012760408, 0012760409, 0012760411, 0012762277, 0012762278, 0012762279, 0012762280, 0012762281, 0012762282, 0012762283, 0012762284, 0012765019, 0012765020, 0012777959, 0012777961, 0012777962, 0012777964, 0012777965, 0012777985, 0012777986, 0012777987, 0012777989, 0012777991, 0012781703, 0012783280, 0012783281, 0012783282, 0012783283, 0012783284, 0012783285, 0012783286, 0012783288, 0012783293, 0012783298, 0012784425, 0012784426, 0012784427, 0012784428, 0012784429, 0012784430, 0012784431, 0012784432, 0012784433, 0012784434, 0012794827, 0012794828, 0012794829, 0012794831, 0012794832, 0012794833, 0012794835, 0012794836, 0012794837, 0012794839, 0012794840, 0012794849, 0012794850, 0012794851, 0012794854, 0012794855, 0012794856, 0012794857, 0012794858, 0012794860, 0012794861, 0012794863, 0012794864, 0012794865, 0012794866, 0012794867, 0012794868, 0012794869, 0012794870, . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic, Inc.
- Distributor
- Worldwide
- Lot
- US: Basic UDI Number 0763000B000262284, GTIN: 00763000871741; 00763000973391; 00763000973506; 00763000993269; 00199150003709. OUS: 00763000871734; 00763000973384; 00763000973513; 00763000811341; 00763000911515; 00763000963675; 00199150003716. Serial Numbers: 227076469, 2301234000, 2301321000, 2301373000, 2301543000, 2301547000, 2301551000, 2301552000, 2301556000, 2301558000, 2301559000, 2301560000, 2301709000, 0012573781, 0012576276, 0012578858, 0012583026, 0012583027, 0012590913, 0012590914, 0012594540, 0012594541, 0012609253, 0012609255, 0012609258, 0012614806, 0012614807, 0012614808, 0012619893, 0012619896, 0012619898, 0012619899, 0012623987, 0012623988, 0012623991, 0012631219, 0012631220, 0012631221, 0012631222, 0012635279, 0012635281, 0012635282, 0012635283, 0012635284, 0012641293, 0012641294, 0012641295, 0012641297, 0012643392, 0012643393, 0012643396, 0012643399, 0012645205, 0012645206, 0012645208, 0012655005, 0012655006, 0012655008, 0012655009, 0012655010, 0012655011, 0012655012, 0012655013, 0012655014, 0012655015, 0012655016, 0012655017, 0012665669, 0012665670, 0012665671, 0012665673, 0012665674, 0012665675, 0012670372, 0012670375, 0012670377, 0012670380, 0012670381, 0012670383, 0012672152, 0012672156, 0012672157, 0012672158, 0012672159, 0012672160, 0012672161, 0012672162, 0012672163, 0012672164, 0012672166, 0012672167, 0012681148, 0012681149, 0012681151, 0012681152, 0012681153, 0012681154, 0012681155, 0012681157, 0012681159, 0012681160, 0012685946, 0012685947, 0012685948, 0012685950, 0012685951, 0012685952, 0012685953, 0012685954, 0012685955, 0012688084, 0012690714, 0012690716, 0012690717, 0012690719, 0012690731, 0012690732, 0012690733, 0012690735, 0012690736, 0012690737, 0012690738, 0012690739, 0012699686, 0012699687, 0012699688, 0012699689, 0012699691, 0012699693, 0012699694, 0012699695, 0012699696, 0012699697, 0012699698, 0012699699, 0012703730, 0012703731, 0012703732, 0012703733, 0012703734, 0012703735, 0012703736, 0012703737, 0012703740, 0012703741, 0012703743, 0012703744, 0012705502, 0012705503, 0012705504, 0012705505, 0012705506, 0012705507, 0012705508, 0012705509, 0012705510, 0012705512, 0012716663, 0012716664, 0012716665, 0012716666, 0012716667, 0012716668, 0012716669, 0012716670, 0012716671, 0012716672, 0012716673, 0012716674, 0012724581, 0012724583, 0012724605, 0012724606, 0012724607, 0012724608, 0012724609, 0012724610, 0012724612, 0012724613, 0012728680, 0012728681, 0012728682, 0012728684, 0012728686, 0012728688, 0012728689, 0012740819, 0012740820, 0012740821, 0012740823, 0012740824, 0012740825, 0012740855, 0012740856, 0012740857, 0012740858, 0012743381, 0012743382, 0012743384, 0012743385, 0012743386, 0012743387, 0012743388, 0012743389, 0012743390, 0012743393, 0012743495, 0012743496, 0012752332, 0012752333, 0012752334, 0012752335, 0012752336, 0012752337, 0012752338, 0012752340, 0012752341, 0012754564, 0012754565, 0012754566, 0012754567, 0012754568, 0012754569, 0012754570, 0012754571, 0012758280, 0012758281, 0012758282, 0012758285, 0012758286, 0012758287, 0012758288, 0012758289, 0012760403, 0012760404, 0012760405, 0012760406, 0012760407, 0012760408, 0012760409, 0012760411, 0012762277, 0012762278, 0012762279, 0012762280, 0012762281, 0012762282, 0012762283, 0012762284, 0012765019, 0012765020, 0012777959, 0012777961, 0012777962, 0012777964, 0012777965, 0012777985, 0012777986, 0012777987, 0012777989, 0012777991, 0012781703, 0012783280, 0012783281, 0012783282, 0012783283, 0012783284, 0012783285, 0012783286, 0012783288, 0012783293, 0012783298, 0012784425, 0012784426, 0012784427, 0012784428, 0012784429, 0012784430, 0012784431, 0012784432, 0012784433, 0012784434, 0012794827, 0012794828, 0012794829, 0012794831, 0012794832, 0012794833, 0012794835, 0012794836, 0012794837, 0012794839, 0012794840, 0012794849, 0012794850, 0012794851, 0012794854, 0012794855, 0012794856, 0012794857, 0012794858, 0012794860, 0012794861, 0012794863, 0012794864, 0012794865, 0012794866, 0012794867, 0012794868, 0012794869, 0012794870,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACareLink SmartSync Device Manager App, Model M01A02, Pulse generator, permanent, implantableNovember 4, 2026
- FDACareLink SmartSync, RF Head Patient Connector, Model 24967, Pulse generator, permanent, implantableNovember 4, 2026
- FDAOmniaSecure Model 3930M, LEAD 3930M74 74 CM DF4 MR 3930M74 LEAD 3930M79 79 CM DF4 MR 3930M79 The MedtronicNovember 4, 2026
- FDACareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantableNovember 4, 2026
- FDAAIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty firstMay 6, 2026
- FDABrand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/CatalogMay 6, 2026
Questions about this recall
Was Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter. recalled?
Yes. FDA published recall Z-1948-2026 on May 6, 2026. Reported hazard: Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit