Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral.
Published August 21, 2013 · Recall initiated: May 9, 2012
Summary
FDA records medical device Class II recall Z-1957-2013, published August 21, 2013 and initiated May 9, 2012. Recalling firm: SpineFrontier, Inc.; Status: Terminated; Reported quantity: 9; Lot: Lot Number: 011836-0001, 011981-001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The nipple of the tipped driver broke off when surgeons were attempting to insert the screw.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- SpineFrontier, Inc.
- Distributor
- Distributed in DC and the states of MA, and OK.
- Lot
- Lot Number: 011836-0001, 011981-001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAA-CIFT SoloFuse Fixed Angle Driver, Model Number 13-32101-01 (Component Part Numbers 14-32107-01 and 14-32106)July 24, 2019
- FDAlnline Persuader- PedFuse Pedicle Screw System; Part Number: 11-80164January 21, 2015
- FDASpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device inMay 14, 2014
- FDAArena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of theJanuary 22, 2014
- FDAStealthStation S7 Framelink Software version 5.4. Aid for locating anatomical structures and planning surgicalAugust 21, 2013
- FDACase Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B,August 21, 2013
Questions about this recall
Was Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral. recalled?
Yes. FDA published recall Z-1957-2013 on August 21, 2013. Reported hazard: The nipple of the tipped driver broke off when surgeons were attempting to insert the screw.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail