Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system.

Published August 21, 2013 · Recall initiated: April 22, 2013

Summary

FDA records medical device Class II recall Z-1958-2013, published August 21, 2013 and initiated April 22, 2013. Recalling firm: SpineFrontier, Inc.; Status: Terminated; Reported quantity: 258 units (30 in distribution at the time of the recall); Lot: Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
SpineFrontier, Inc.
Distributor
Worldwide Distribution-USA including the states of FL, OH, MA, and TX, and the countries of Mexico, and Dominican Republic.
Lot
Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system. recalled?

Yes. FDA published recall Z-1958-2013 on August 21, 2013. Reported hazard: Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail