Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system.
Published August 21, 2013 · Recall initiated: April 22, 2013
Summary
FDA records medical device Class II recall Z-1958-2013, published August 21, 2013 and initiated April 22, 2013. Recalling firm: SpineFrontier, Inc.; Status: Terminated; Reported quantity: 258 units (30 in distribution at the time of the recall); Lot: Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- SpineFrontier, Inc.
- Distributor
- Worldwide Distribution-USA including the states of FL, OH, MA, and TX, and the countries of Mexico, and Dominican Republic.
- Lot
- Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAA-CIFT SoloFuse Fixed Angle Driver, Model Number 13-32101-01 (Component Part Numbers 14-32107-01 and 14-32106)July 24, 2019
- FDAlnline Persuader- PedFuse Pedicle Screw System; Part Number: 11-80164January 21, 2015
- FDASpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device inMay 14, 2014
- FDAArena-L Trial Handle; Part: 11-81002 The trial heads are smooth plates that mimic the footprint of theJanuary 22, 2014
- FDAInvue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axisAugust 21, 2013
- FDADisposable Temperature Probes, (sterile), Sensor diameter: 3mm/0.12 in; 25 units per box. Vital Signs Inc. aAugust 21, 2013
Questions about this recall
Was Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system. recalled?
Yes. FDA published recall Z-1958-2013 on August 21, 2013. Reported hazard: Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail