epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.

Published August 21, 2013 · Recall initiated: March 13, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Epocal
Distributor
Worldwide Distribution-USA (nationwide) including the District of Columbia, US Virgin Islands and Puerto Rico, and the states of CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MS, NC, ND, NE, NH, NJ, NM, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VE, WA, WI, and WV and the coutries of India, Argentina, Australia, Azerbaijan, Belarus, Canada, China, Colombia, Egypt, Germany, Guatemala, Israel, Japan, Malaysia, Mexico, Nepal, Netherlands, Peru, Philippines, Russia Fed., Singapore, South Africa, South korea, Thailand, and Vietnam.
Lot
LOT #'s EXP. DATES 07-12283-00 03/26/2013 07-12284-00 03/27/2013 07-12285-00 03/28/2013 07-12289-00 04/01/2013 07-12290-00 04/02/2013 07-12292-00 04/04/2013 07-12293-00 04/05/2013 07-12296-00 04/08/2013 07-12297-00 04/09/2013 07-12298-00 04/10/2013 07-12300-00 04/12/2013 07-12303-00 04/15/2013 07-12305-00 04/17/2013 07-12306-00 04/18/2013 07-12307-00 04/19/2013 07-12310-00 04/22/2013 07-12312-00 04/24/2013 07-12313-00 04/25/2013 07-12314-00 04/26/2013 07-12317-00 04/29/2013 07-12320-00 05/02/2013 07-12321-00 05/03/2013 07-12324-00 05/06/2013 07-12325-00 05/07/2013 07-12327-00 05/09/2013 07-12331-00 05/13/2013 07-12333-00 05/15/2013 07-12335-00 05/17/2013 07-12338-00 05/20/2013 07-12340-00 05/22/2013 07-12341-00 05/23/2013 07-12345-00 05/27/2013 07-12346-00 05/28/2013 07-12347-00 05/29/2013 07-12349-00 05/31/2013 07-12352-00 06/03/2013 07-12354-00 06/05/2013 07-12355-00 06/06/2013 07-12362-00 06/13/2013 07-12363-00 06/14/2013 07-13002-00 06/19/2013 07-13003-00 06/20/2013 07-13004-00 06/21/2013 07-13007-00 06/24/2013 07-13008-00 06/25/2013 07-13009-00 06/26/2013 07-13010-00 06/27/2013 07-13014-00 07/01/2013 07-13015-00 07/02/2013 07-13017-00 07/04/2013
Risk classification
Class II
Area (FDA)
device
Country
Canada