ADVIA Centaur® Systems TnI-Ultra assay; Catalog Number 02789602 (100 tests), SMN 10317708 and 02790309 (500 tests), SMN 10317790

Published July 9, 2014 · Recall initiated: May 30, 2014

Summary

FDA records medical device Class II recall Z-1991-2014, published July 9, 2014 and initiated May 30, 2014. Recalling firm: Siemens Healthcare Diagnostics, Inc; Status: Terminated; Reported quantity: 34,283 kits; Lot: Kit Lots Ending In: 078, 079, 082, 083, 084. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The solid phase reagent in some of the ADVIA Centaur® TnI -Ultra ReadyPacks®, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc
Distributor
Worldwide Distribution-USA (nationwide) including PR and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Canada, Mexico, Argentina, Austria, Australia. Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Great Britain, Greece, Guadeloupe, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Libya, Macedonia, Malaysia, ,Malta, Martinique, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Republic of Korea, Romania, Russian Federation, Serbia, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sri Lanka, South Africa, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Vietnam.
Lot
Kit Lots Ending In: 078, 079, 082, 083, 084
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ADVIA Centaur® Systems TnI-Ultra assay; Catalog Number 02789602 (100 tests), SMN 10317708 and 02790309 (500 tests), SMN 10317790 recalled?

Yes. FDA published recall Z-1991-2014 on July 9, 2014. Reported hazard: The solid phase reagent in some of the ADVIA Centaur® TnI -Ultra ReadyPacks®, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter