Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
Published August 28, 2013 · Recall initiated: July 17, 2013
Summary
FDA records medical device Class II recall Z-1997-2013, published August 28, 2013 and initiated July 17, 2013. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 68,116 units; Lot: Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/31/2014 M208454 21Sep2012 02/28/2014 M210079 19Oct2012 03/31/2014 M211100 30Nov2012 05/31/2014 M212122 20Dec2012 06/30/2014 M212203 17Jan2013 06/30/2014 M301418 22Feb2013 & 28Feb2013 07/31/2014 M202086 24Feb2012 07/31/2013 M203081 16Mar2012 07/31/2013 M203081A 23Mar2012 07/31/2013 M203262 06Apr2012 08/31/2013 M203098 13Apr2012 08/31/2013 M203098A 20Apr2012 08/31/2013 M203098B 25Apr2012 08/31/2013 M203080 27Apr2012 10/31/2013 M203080A 04May2012 10/31/2013 M203080B 11May2012 10/31/2013 M204276 18May2012 09/30/2013 M205199 12May2012 09/30/2013 M205288 22May2012 11/30/2013 M206021 06June2012 11/30/2013 M206166 18June2012 12/31/2013 M207072 05July2012 12/31/2013 M207143 27July2012 01/31/2014 M206259 26June2012 12/31/2013 M207332 30July2012 01/31/2014 M208345 22Aug2012 01/31/2014 M208272 24Aug2012 02/28/2014 M208454 06Sep2012 02/28/2014 M209135 19Sep2012 03/31/2014 M210079 04Oct2012 03/31/2014 M210169 27Oct2012 04/30/2014 M211100 10Nov2012 05/31/2014 M211078 01Dec2012 05/31/2014 M212122 12Dec2012 06/30/2014 M212203 03Jan2013 06/30/2014 M301324 16Jan2013 07/31/2014 M301418 11Feb2013 07/31/2014 M302410 27Feb2013 08/31/2014 M303157 15Mar2013 09/30/2014 M303223 26Mar2013 09/30/2014 M304130 09Apr2013 09/30/2014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide Distribution - USA and the country of Canada.
- Lot
- Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/31/2014 M208454 21Sep2012 02/28/2014 M210079 19Oct2012 03/31/2014 M211100 30Nov2012 05/31/2014 M212122 20Dec2012 06/30/2014 M212203 17Jan2013 06/30/2014 M301418 22Feb2013 & 28Feb2013 07/31/2014 M202086 24Feb2012 07/31/2013 M203081 16Mar2012 07/31/2013 M203081A 23Mar2012 07/31/2013 M203262 06Apr2012 08/31/2013 M203098 13Apr2012 08/31/2013 M203098A 20Apr2012 08/31/2013 M203098B 25Apr2012 08/31/2013 M203080 27Apr2012 10/31/2013 M203080A 04May2012 10/31/2013 M203080B 11May2012 10/31/2013 M204276 18May2012 09/30/2013 M205199 12May2012 09/30/2013 M205288 22May2012 11/30/2013 M206021 06June2012 11/30/2013 M206166 18June2012 12/31/2013 M207072 05July2012 12/31/2013 M207143 27July2012 01/31/2014 M206259 26June2012 12/31/2013 M207332 30July2012 01/31/2014 M208345 22Aug2012 01/31/2014 M208272 24Aug2012 02/28/2014 M208454 06Sep2012 02/28/2014 M209135 19Sep2012 03/31/2014 M210079 04Oct2012 03/31/2014 M210169 27Oct2012 04/30/2014 M211100 10Nov2012 05/31/2014 M211078 01Dec2012 05/31/2014 M212122 12Dec2012 06/30/2014 M212203 03Jan2013 06/30/2014 M301324 16Jan2013 07/31/2014 M301418 11Feb2013 07/31/2014 M302410 27Feb2013 08/31/2014 M303157 15Mar2013 09/30/2014 M303223 26Mar2013 09/30/2014 M304130 09Apr2013 09/30/2014
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDAGENESIS(R) II CRUCIATE RETAINING ARTICULAR INSERT, TRIAL, SIZE 3-4, A/P 48 MM, M/L 68 MM, QTY 1, NON-STERILE,August 28, 2013
- FDA00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR,August 28, 2013
Questions about this recall
Was Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples. recalled?
Yes. FDA published recall Z-1997-2013 on August 28, 2013. Reported hazard: Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
What should I do if I have this product?
Voluntary: Firm initiated · Letter