Heraeus, PALAMIX uno. Material Number: 66057893.
Published May 6, 2026 · Recall initiated: March 25, 2026
Summary
FDA records medical device Class III recall Z-1997-2026, published May 6, 2026 and initiated March 25, 2026. Recalling firm: Heraeus Medical GmbH (Dental Division); Status: Ongoing; Reported quantity: 51,771 units; Lot: Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060. Class III means the product is not likely to cause adverse health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Heraeus Medical GmbH (Dental Division)
- Distributor
- Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
- Lot
- Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAPALACOS MV+G pro 40; PALACOS MV+G pro 80. Model/Catalog Number: 5081290; 5081291. Radiopaque, poly(methylOctober 7, 2026
- FDAPALACOS R pro 40; PALACOS R pro 80. Model/Catalog Numbers: 5081286; 5081287. Radiopaque, poly (methylOctober 7, 2026
- FDAPALACOS MV pro 40; PALACOS MV pro 80. Model/Catalog Number: 5150633; 5150635. Radiopaque, poly(methylOctober 7, 2026
- FDAPALACOS R+G pro 40; PALACOS R+G pro 80. Model/Catalog Number: 5081288; 5081289. Radiopaque, poly(methylOctober 7, 2026
- FDAiLet Bionic Pancreas, REF: BB1001May 6, 2026
- FDAHeraeus, PALAMIX duo. Material Number: 66057897.May 6, 2026
Questions about this recall
Was Heraeus, PALAMIX uno. Material Number: 66057893. recalled?
Yes. FDA published recall Z-1997-2026 on May 6, 2026. Reported hazard: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
What should I do if I have this product?
Voluntary: Firm initiated · Letter