Heraeus, PALAMIX duo. Material Number: 66057897.

Published May 6, 2026 · Recall initiated: March 25, 2026

Summary

FDA records medical device Class III recall Z-1998-2026, published May 6, 2026 and initiated March 25, 2026. Recalling firm: Heraeus Medical GmbH (Dental Division); Status: Ongoing; Reported quantity: 1,125 units; Lot: Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215. Class III means the product is not likely to cause adverse health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Heraeus Medical GmbH (Dental Division)
Distributor
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Lot
Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
Risk classification
Class III
Area (FDA)
device
Country
Germany

Related recalls

Questions about this recall

Was Heraeus, PALAMIX duo. Material Number: 66057897. recalled?

Yes. FDA published recall Z-1998-2026 on May 6, 2026. Reported hazard: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

What should I do if I have this product?

Voluntary: Firm initiated · Letter