PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application
Published July 5, 2023 · Recall initiated: May 10, 2023
Summary
FDA records medical device Class II recall Z-2004-2023, published July 5, 2023 and initiated May 10, 2023. Recalling firm: Permobil; Status: Ongoing; Reported quantity: 18,905 units; Lot: All units with software versions 1.0.00 to 1.0.04. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Permobil
- Distributor
- Worldwide distribution.
- Lot
- All units with software versions 1.0.00 to 1.0.04
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATizanidine Tablets, USP, 4 mg, 100 Tablets (10 tablets x 10 unit dose blister packs) per carton, Rx Only,July 12, 2023
- FDATizanidine Hydrochloride Tablet 4mg, packaged in a) 20 count-bottles (NDC 68788-7781-2), b)30-count bottlesJuly 12, 2023
- FDADesign Options® / Epidural anesthesia kit (10 count carton)July 5, 2023
- FDAImmunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor theJuly 5, 2023
- FDAMEGADYNE MEGA SOFT Universal Plus Patient Return Electrode. used during electrosurgeryJuly 5, 2023
- FDAEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080060July 5, 2023
Questions about this recall
Was PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application recalled?
Yes. FDA published recall Z-2004-2023 on July 5, 2023. Reported hazard: When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit