Back to search
FDAFDA · Food and drugs·Recall number Z-2021-2023

Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes. Catalog Number: IA2640

Published July 5, 2023 · Recall initiated: May 4, 2023

Summary

FDA records medical device Class III recall Z-2021-2023, published July 5, 2023 and initiated May 4, 2023. Recalling firm: Randox Laboratories Ltd.; Status: Ongoing; Reported quantity: 1 kit; Lot: GTIN:05055273203868 Batch/Lot Number: 2031EC Exp. Date: 28th Nov 23. Class III means the product is not likely to cause adverse health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Randox Laboratories Ltd.
Distributor
CA, IL, VA, WA, WI including PR
Lot
GTIN:05055273203868 Batch/Lot Number: 2031EC Exp. Date: 28th Nov 23
Risk classification
Class III
Area (FDA)
device
Country
Ireland

Related recalls

Questions about this recall

Was Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes. Catalog Number: IA2640 recalled?

Yes. FDA published recall Z-2021-2023 on July 5, 2023. Reported hazard: Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch 583135 lot 2031EC has been reassigned on the Roche Cobas e801 due to High recovery outside range

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail