K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90109, UDI # 10888857108226, Size 24x6mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.
Published May 27, 2020 · Recall initiated: April 13, 2020
Summary
FDA records medical device Class II recall Z-2009-2020, published May 27, 2020 and initiated April 13, 2020. Recalling firm: K2M, Inc; Status: Terminated; Reported quantity: N/A; Lot: Lot # FKWW. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- K2M, Inc
- Distributor
- Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.
- Lot
- Lot # FKWW
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEverest MI XT Inner Dilator, Catalog Number 5101-90167March 9, 2022
- FDAYukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (theDecember 9, 2020
- FDAK2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, CatalogMay 27, 2020
- FDAWillamette Valley Pie Company Cherry Pie, packaged in clamshells container, six containers per case. Each pieMay 27, 2020
- FDAWillamette Valley Pie Company Raspberry Rhubarb Pie, packaged in clamshells container, six containers perMay 27, 2020
- FDASTEALTH SPRING CLIP 6mm latis (3/4 Force). REF/UDI::A1603/(01)00607915110581 - Product Usage: The device isMay 27, 2020
Questions about this recall
Was K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90109, UDI # 10888857108226, Size 24x6mm, 6 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy. recalled?
Yes. FDA published recall Z-2009-2020 on May 27, 2020. Reported hazard: The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter