Back to search
FDAFDA · Food and drugs·Recall number Z-2022-2014

STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches

Published July 16, 2014 · Recall initiated: May 29, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
B. Braun Medical, Inc.
Distributor
USA (nationwide) including the states of PA, IL, NC, FL, IL, CA, CO, NE, LA, NC, LA, VA, TN, NE, TX, and KS.
Lot
Product codes: 473547 and 473549 lot #s and expiration dates: 61356774, 1/31/2019; 61362114, 2/28/2019; and 61367665, 3/31/2019.
Risk classification
Class II
Area (FDA)
device
Country
United States