The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures.
Published July 16, 2014 · Recall initiated: June 27, 2014
Summary
FDA records medical device Class II recall Z-2026-2014, published July 16, 2014 and initiated June 27, 2014. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 3689; Lot: All Model Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide Distribution - US (nationwide in the states of AL, AK, AZ, AR, CA,CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY in the countries of Canada, India, Vietnam, China, India, Korea, Mexico, costa Rica, Brazil, Taiwan, Australia, New Zealand, El Salvador, Indonesia, Hong Kong, Singapore, Nepal.
- Lot
- All Model Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDASmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894July 16, 2014
- FDAThe Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication ManagementJuly 16, 2014
Questions about this recall
Was The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures. recalled?
Yes. FDA published recall Z-2026-2014 on July 16, 2014. Reported hazard: STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.
What should I do if I have this product?
Voluntary: Firm initiated · Letter