FDAFDA · Food and drugs·Recall number Z-2029-2024

OES 4000 Hysteroscope, Model/Catalog Number: A4674A

Published June 12, 2024 · Recall initiated: April 18, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
US Nationwide distribution in the state of NY.
Lot
Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;
Risk classification
Class II