OES 4000 Hysteroscope, Model/Catalog Number: A4674A
Published June 12, 2024 · Recall initiated: April 18, 2024
Summary
FDA records medical device Class II recall Z-2029-2024, published June 12, 2024 and initiated April 18, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 1 unit; Lot: Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- US Nationwide distribution in the state of NY.
- Lot
- Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDAPhenazopyridine HCl Tablets USP, 100 mg, 100-count bottles, Rx Only, Manufactured by: Winder Laboratories,June 19, 2024
- FDAPhilips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator ManualJune 12, 2024
Questions about this recall
Was OES 4000 Hysteroscope, Model/Catalog Number: A4674A recalled?
Yes. FDA published recall Z-2029-2024 on June 12, 2024. Reported hazard: The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
What should I do if I have this product?
Voluntary: Firm initiated · Letter