OES 4000 Hysteroscope, Model/Catalog Number: A4674A

Published June 12, 2024 · Recall initiated: April 18, 2024

Summary

FDA records medical device Class II recall Z-2029-2024, published June 12, 2024 and initiated April 18, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 1 unit; Lot: Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
US Nationwide distribution in the state of NY.
Lot
Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OES 4000 Hysteroscope, Model/Catalog Number: A4674A recalled?

Yes. FDA published recall Z-2029-2024 on June 12, 2024. Reported hazard: The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.

What should I do if I have this product?

Voluntary: Firm initiated · Letter