MOSAIQ Oncology Information System
Published July 6, 2016 · Recall initiated: June 17, 2016
Summary
FDA records medical device Class II recall Z-2052-2016, published July 6, 2016 and initiated June 17, 2016. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 69; Lot: Versions 2.62 and higher. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It is possible that that the selected Plan-of-the Day could be accidently changed by scrolling the mouse wheel up and down and inadvertently selecting another plan.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- US Distribution including Puerto Rico and to the states of : AK, AZ, CA, FL, GA, IL, IN, LA, MD, MI, MO, NJ, NY, OH, OR, PA, TN, TX, UT, VA, WA, and to the countries of : Canada, Denmark, Finland, Germany, Israel, Japan, Netherlands, Norway, Sweden, United Kingdom
- Lot
- Versions 2.62 and higher
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended forJuly 15, 2026
- FDABrand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: TheJuly 15, 2026
- FDABrand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading ApplicatorJuly 15, 2026
- FDAMOSAIQ Oncology Information SystemJune 17, 2026
- FDAPiggly Wiggly, Yellow Cut Corn, Net Wt 32oz (2 lbs), 907g bags, Distributed by Piggly Wiggly LLC, Keene, NHJuly 6, 2016
- FDASymbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is aJuly 6, 2016
Questions about this recall
Was MOSAIQ Oncology Information System recalled?
Yes. FDA published recall Z-2052-2016 on July 6, 2016. Reported hazard: It is possible that that the selected Plan-of-the Day could be accidently changed by scrolling the mouse wheel up and down and inadvertently selecting another plan.
What should I do if I have this product?
Voluntary: Firm initiated · Letter