Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
Published September 4, 2013 · Recall initiated: May 31, 2013
Summary
FDA records medical device Class II recall Z-2057-2013, published September 4, 2013 and initiated May 31, 2013. Recalling firm: Synthes (USA) Products LLC; Status: Terminated; Reported quantity: 119, 891; Lot: Part number J8301-F. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes (USA) Products LLC
- Distributor
- Nationwide Distribution including the states of AK. AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MD, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI SC, SD, TN, TX, UT, VA, VT, WA WI and WV.
- Lot
- Part number J8301-F
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASTARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screwJune 5, 2024
- FDA13 mm Titanium Hindfoot Arthrodesis Cannulated Nails EX, sterile -Intended to facilitate tibiotalocalcanealMarch 13, 2024
- FDARAPIDSORB IPS Battery Pack, Sterile-Intended for use in non-load bearing fracture repair and reconstructiveMarch 13, 2024
- FDAOpal, Intervertebral Cage, 10x32mm, h 11mm, Revolvable-Indicated for degenerative spine disease Part Number:March 13, 2024
- FDAIntegra Leyla Ball Joint Clamp, also as included in Leyla Table Mounting Hardware. The Leyla RetractionSeptember 4, 2013
- FDASynthes Trauma Nail System. The devices are indication for bone fixation.September 4, 2013
Questions about this recall
Was Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones. recalled?
Yes. FDA published recall Z-2057-2013 on September 4, 2013. Reported hazard: The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm
What should I do if I have this product?
Voluntary: Firm initiated · Letter