Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits.

Published July 23, 2014 · Recall initiated: June 18, 2014

Summary

FDA records medical device Class II recall Z-2077-2014, published July 23, 2014 and initiated June 18, 2014. Recalling firm: Spacelabs Healthcare Inc; Status: Terminated; Reported quantity: 8 units total (3 units in US); Lot: Serial Numbers: 2518-000183 2518-000184 2518-000185 2518-000186 2518-000206 2518-000207 2518-000208 2518-000209. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Spacelabs Healthcare Inc
Distributor
Worldwide Distribution - US in the state of Georgia and in the countries of : Malaysia, Paraguay, and France.
Lot
Serial Numbers: 2518-000183 2518-000184 2518-000185 2518-000186 2518-000206 2518-000207 2518-000208 2518-000209
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits. recalled?

Yes. FDA published recall Z-2077-2014 on July 23, 2014. Reported hazard: With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.

What should I do if I have this product?

Voluntary: Firm initiated · Letter