Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits.
Published July 23, 2014 · Recall initiated: June 18, 2014
Summary
FDA records medical device Class II recall Z-2077-2014, published July 23, 2014 and initiated June 18, 2014. Recalling firm: Spacelabs Healthcare Inc; Status: Terminated; Reported quantity: 8 units total (3 units in US); Lot: Serial Numbers: 2518-000183 2518-000184 2518-000185 2518-000186 2518-000206 2518-000207 2518-000208 2518-000209. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Spacelabs Healthcare Inc
- Distributor
- Worldwide Distribution - US in the state of Georgia and in the countries of : Malaysia, Paraguay, and France.
- Lot
- Serial Numbers: 2518-000183 2518-000184 2518-000185 2518-000186 2518-000206 2518-000207 2518-000208 2518-000209
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA91496 Ultraview SL Command Module, Options A, B, C, and I. Option A: ECG/Respiration, SpO2, Temperature x2 &August 10, 2016
- FDAThe Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280 is used to transmit patient(s)April 27, 2016
- FDASpacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. TheOctober 1, 2014
- FDAARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in aSeptember 10, 2014
- FDATritium Sternal Plate System, Screw, 02.7mm X 12mm, 12 Pack, Sterile. The Pioneer Surgical Cable Plate SystemJuly 23, 2014
- FDATitan T 5K Motor Product Usage: Low-speed motor is used with various attachments to perform a variety ofJuly 23, 2014
Questions about this recall
Was Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits. recalled?
Yes. FDA published recall Z-2077-2014 on July 23, 2014. Reported hazard: With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.
What should I do if I have this product?
Voluntary: Firm initiated · Letter