The Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280 is used to transmit patient(s) physiological measurements and data to a monitor.

Published April 27, 2016 · Recall initiated: March 28, 2016

Summary

FDA records medical device Class II recall Z-1502-2016, published April 27, 2016 and initiated March 28, 2016. Recalling firm: Spacelabs Healthcare Inc; Status: Terminated; Reported quantity: 217 units of XTRs and 531 units of PCBAs (164 XTRs and 425 PCBAs in US and 53 XTRs and 106 PCBAs outside the US); Lot: Affected PCBAs have no serial numbers. Below is the Part Numbers: 040-1736-00 KIT,UPGRADE,PCBA,QUAD RCVR,T BAND, 96280; 050-0758-00 REPLACEMENT KIT, PCBA,QUAD RCVR, Q BAND, 96280; 050-0759-00 REPLACEMENT KIT, PCBA,QUAD RCVR, T BAND, 96280; 050-0760-00 REPLACEMENT KIT, PCBA,QUAD RCVR, V BAND, 96280; 050-0758-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, Q BAND, 96280 EXCHANGE; 050-0759-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, T BAND, 96280 EXCHANGE; 050-0760-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, V BAND, 96280 EXCHANGE; 050-0519-00 CM KIT,ET,600 MHZ-1400MHZ,NA, 96280; 050-0532-00 CM KIT,ET,400MHZ-600MHZ,INTERNATIONAL, 96280. **************************SERIAL NUMBERS OF XTRs IN THE US************ 6280-000051, 6280-000053, 6280-000056, 6280-000057, 6280-000058, 6280-000059, 6280-000066, 6280-000067, 6280-000068, 6280-000077, 6280-000078, 6280-000079, 6280-000080, 6280-000083, 6280-000084, 6280-000085, 6280-000086, 6280-000087, 6280-000088, 6280-000090, 6280-000092, 6280-000093, 6280-000095, 6280-000096, 6280-000097, 6280-000098, 6280-000099, 6280-000100, 6280-000101, 6280-000102, 6280-000103, 6280-000104, 6280-000105, 6280-000106, 6280-000107, 6280-000108, 6280-000109, 6280-000110, 6280-000111, 6280-000112, 6280-000113, 6280-000114, 6280-000115, 6280-000116, 6280-000117, 6280-000118, 6280-000119, 6280-000120, 6280-000121, 6280-000122, 6280-000123, 6280-000124, 6280-000125, 6280-000126, 6280-000127, 6280-000128, 6280-000129, 6280-000131, 6280-000132, 6280-000133, 6280-000135, 6280-000136, 6280-000137, 6280-000138, 6280-000140, 6280-000141, 6280-000142, 6280-000143, 6280-000145, 6280-000146, 6280-000147, 6280-000148, 6280-000149, 6280-000151, 6280-000152, 6280-000153, 6280-000154, 6280-000155, 6280-000156, 6280-000157, 6280-000158, 6280-000159, 6280-000160, 6280-000161, 6280-000167, 6280-000168, 6280-000169, 6280-000170, 6280-000171, 6280-000172, 6280-000173, 6280-000174, 6280-000175, 6280-000176, 6280-000177, 6280-000200, 6280-000227, 6280-000228, 6280-000229, 6280-000233, 6280-000234, 6280-000235, 6280-000236, 6280-000238, 6280-000239, 6280-000240, 6280-000241, 6280-000242, 6280-000243, 6280-000245, 6280-000248, 6280-000250, 6280-000251, 6280-000252, 6280-000253, 6280-000254, 6280-000255, 6280-000256, 6280-000257, 6280-000259, 6280-000260, 6280-000261, 6280-000263, 6280-000264, 6280-000265, 6280-000266, 6280-000267, 6280-000268, 6280-000269, 6280-000271, 6280-000272, 6280-000273, 6280-000274, 6280-000275, 6280-000278, 6280-000279, 6280-000280, 6280-000281, 6280-000283, 6280-000284, 6280-000285, 6280-000286, 6280-000287, 6280-000288, 6280-000289, 6280-000298, 6280-000299, 6280-000301, 6280-000302, 6280-000303, 6280-000304, 6280-000306, 6280-000307, 6280-000308, 6280-000309, 6280-000310, 6280-000311, 6280-000312, 6280-000313, 6280-000314, 6280-000315, 6280-000316, 6280-000319, and 6281-000144. *********SERIAL NUMBERS OF XTRs OUTSIDE THE US************ 6280-000048, 6280-000050, 6280-000052, 6280-000060, 6280-000061, 6280-000062, 6280-000063, 6280-000064, 6280-000065, 6280-000076, 6280-000094, 6280-000178, 6280-000179, 6280-000181, 6280-000182, 6280-000183, 6280-000184, 6280-000185, 6280-000186, 6280-000188, 6280-000189, 6280-000190, 6280-000191, 6280-000193, 6280-000194, 6280-000195, 6280-000196, 6280-000197, 6280-000198, 6280-000199, 6280-000202, 6280-000225, 6280-000226, 6280-000231, 6280-000232, 6280-000237, 6280-000246, 6280-000247, 6280-000249, 6280-000262, 6280-000270, 6280-000276, 6280-000282, 6280-000290, 6280-000291, 6280-000292, 6280-000293, 6280-000294, 6280-000295, 6280-000297, 6280-000300, 6280-000305, and 6280-000317. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports of an abnormally high frequency of squelch events on Xhibit Telemetry Receivers (XTR), which is indicative of the loss of the telemetry data signal. In some cases, the Xhibit Central Monitor was also reported to display an Xhibit Offline error message for all patients being monitored.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Spacelabs Healthcare Inc
Distributor
distributed in AL, AZ, CA, CO, CT, FL, ID, IN, KY, MI, MN, MO, MS, MT, SC, TX, WY and in the following countries: AUSTRALIA, CANADA, CHINA, FRANCE, NETHERLANDS, PHILIPPINES, ROMANIA, and UNITED KINGDOM.
Lot
Affected PCBAs have no serial numbers. Below is the Part Numbers: 040-1736-00 KIT,UPGRADE,PCBA,QUAD RCVR,T BAND, 96280; 050-0758-00 REPLACEMENT KIT, PCBA,QUAD RCVR, Q BAND, 96280; 050-0759-00 REPLACEMENT KIT, PCBA,QUAD RCVR, T BAND, 96280; 050-0760-00 REPLACEMENT KIT, PCBA,QUAD RCVR, V BAND, 96280; 050-0758-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, Q BAND, 96280 EXCHANGE; 050-0759-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, T BAND, 96280 EXCHANGE; 050-0760-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, V BAND, 96280 EXCHANGE; 050-0519-00 CM KIT,ET,600 MHZ-1400MHZ,NA, 96280; 050-0532-00 CM KIT,ET,400MHZ-600MHZ,INTERNATIONAL, 96280. **************************SERIAL NUMBERS OF XTRs IN THE US************ 6280-000051, 6280-000053, 6280-000056, 6280-000057, 6280-000058, 6280-000059, 6280-000066, 6280-000067, 6280-000068, 6280-000077, 6280-000078, 6280-000079, 6280-000080, 6280-000083, 6280-000084, 6280-000085, 6280-000086, 6280-000087, 6280-000088, 6280-000090, 6280-000092, 6280-000093, 6280-000095, 6280-000096, 6280-000097, 6280-000098, 6280-000099, 6280-000100, 6280-000101, 6280-000102, 6280-000103, 6280-000104, 6280-000105, 6280-000106, 6280-000107, 6280-000108, 6280-000109, 6280-000110, 6280-000111, 6280-000112, 6280-000113, 6280-000114, 6280-000115, 6280-000116, 6280-000117, 6280-000118, 6280-000119, 6280-000120, 6280-000121, 6280-000122, 6280-000123, 6280-000124, 6280-000125, 6280-000126, 6280-000127, 6280-000128, 6280-000129, 6280-000131, 6280-000132, 6280-000133, 6280-000135, 6280-000136, 6280-000137, 6280-000138, 6280-000140, 6280-000141, 6280-000142, 6280-000143, 6280-000145, 6280-000146, 6280-000147, 6280-000148, 6280-000149, 6280-000151, 6280-000152, 6280-000153, 6280-000154, 6280-000155, 6280-000156, 6280-000157, 6280-000158, 6280-000159, 6280-000160, 6280-000161, 6280-000167, 6280-000168, 6280-000169, 6280-000170, 6280-000171, 6280-000172, 6280-000173, 6280-000174, 6280-000175, 6280-000176, 6280-000177, 6280-000200, 6280-000227, 6280-000228, 6280-000229, 6280-000233, 6280-000234, 6280-000235, 6280-000236, 6280-000238, 6280-000239, 6280-000240, 6280-000241, 6280-000242, 6280-000243, 6280-000245, 6280-000248, 6280-000250, 6280-000251, 6280-000252, 6280-000253, 6280-000254, 6280-000255, 6280-000256, 6280-000257, 6280-000259, 6280-000260, 6280-000261, 6280-000263, 6280-000264, 6280-000265, 6280-000266, 6280-000267, 6280-000268, 6280-000269, 6280-000271, 6280-000272, 6280-000273, 6280-000274, 6280-000275, 6280-000278, 6280-000279, 6280-000280, 6280-000281, 6280-000283, 6280-000284, 6280-000285, 6280-000286, 6280-000287, 6280-000288, 6280-000289, 6280-000298, 6280-000299, 6280-000301, 6280-000302, 6280-000303, 6280-000304, 6280-000306, 6280-000307, 6280-000308, 6280-000309, 6280-000310, 6280-000311, 6280-000312, 6280-000313, 6280-000314, 6280-000315, 6280-000316, 6280-000319, and 6281-000144. *********SERIAL NUMBERS OF XTRs OUTSIDE THE US************ 6280-000048, 6280-000050, 6280-000052, 6280-000060, 6280-000061, 6280-000062, 6280-000063, 6280-000064, 6280-000065, 6280-000076, 6280-000094, 6280-000178, 6280-000179, 6280-000181, 6280-000182, 6280-000183, 6280-000184, 6280-000185, 6280-000186, 6280-000188, 6280-000189, 6280-000190, 6280-000191, 6280-000193, 6280-000194, 6280-000195, 6280-000196, 6280-000197, 6280-000198, 6280-000199, 6280-000202, 6280-000225, 6280-000226, 6280-000231, 6280-000232, 6280-000237, 6280-000246, 6280-000247, 6280-000249, 6280-000262, 6280-000270, 6280-000276, 6280-000282, 6280-000290, 6280-000291, 6280-000292, 6280-000293, 6280-000294, 6280-000295, 6280-000297, 6280-000300, 6280-000305, and 6280-000317.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280 is used to transmit patient(s) physiological measurements and data to a monitor. recalled?

Yes. FDA published recall Z-1502-2016 on April 27, 2016. Reported hazard: Reports of an abnormally high frequency of squelch events on Xhibit Telemetry Receivers (XTR), which is indicative of the loss of the telemetry data signal. In some cases, the Xhibit Central Monitor was also reported to display an Xhibit Offline error message for all patients being monitored.

What should I do if I have this product?

Voluntary: Firm initiated · Letter