Synthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System). Orthopedic surgical instrument.
Published July 30, 2014 · Recall initiated: May 29, 2014
Summary
FDA records medical device Class II recall Z-2087-2014, published July 30, 2014 and initiated May 29, 2014. Recalling firm: Synthes, Inc.; Status: Terminated; Reported quantity: 180; Lot: part no. 03.111.600, lot no. 10-5147, and part no. 03.111.601, lot no. 10-5147. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System) were found to be labeled incorrectly. Part no. 03.111.600 (Right) was found in the package labeled part no. 03.111.601 (Left). The laser etching denoting the orientation on the part is correct.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes, Inc.
- Distributor
- Distribution US Nationwide and Bermuda.
- Lot
- part no. 03.111.600, lot no. 10-5147, and part no. 03.111.601, lot no. 10-5147
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated forJune 24, 2015
- FDAPull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling isJune 24, 2015
- FDANOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 intendedJuly 30, 2014
- FDAHemochron Jr. Whole Blood Microcoagulation System APTT Cuvette. For in vitro diagnostic coagulation test.July 30, 2014
Questions about this recall
Was Synthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System). Orthopedic surgical instrument. recalled?
Yes. FDA published recall Z-2087-2014 on July 30, 2014. Reported hazard: The Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System) were found to be labeled incorrectly. Part no. 03.111.600 (Right) was found in the package labeled part no. 03.111.601 (Left). The laser etching denoting the orientation on the part is correct.
What should I do if I have this product?
Voluntary: Firm initiated · Letter