GE Ultrasound ViewPoint 6. Product Usage: ViewPoint is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images and generation of medical reports primarily for diagnostic ultrasound. ViewPoint provides the user the ability to including images, drawings, and charts into medical reports. ViewPoint is designed to accept, transfer, display, calculate, store and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint displays to the user, can be used for diagnostic purposes. ViewPoint is intended for professional use only. ViewPoint is not intended to be used as an automated diagnosis system.
Published September 4, 2013 · Recall initiated: August 7, 2013
Summary
FDA records medical device Class II recall Z-2098-2013, published September 4, 2013 and initiated August 7, 2013. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 225; Lot: software versions 6.2, 6.2.1, 6.3, and 6.3.1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE has recently become aware of a potential safety issue due to the incorrect calculation of the pulsatility index value (PIV) for venous flow measurements associated with obstetric reporting of the GE ViewPoint 6 system. The ViewPoint calculated Doppler PIV for veins could be incorrect. An incorrect calculation of the ductus venosus PIV could lead to false high risk for trisomy 13, 18, and 21,
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare, LLC
- Distributor
- Worldwide Distribution - USA Nationwide including Puerto Rico and the countries of: ANDORRA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BRAZIL, BULGARIA, CANADA, CHILE, COLOMBIA, CZECH REPUBLIC, ESTONIA, FINLAND, GERMANY, HONG KONG, INDIA, INDONESIA, IRAQ, ITALY, JAPAN, KUWAIT, MALAYSIA, Norway, OMAN, PALESTINIAN TERRITORY, OCCUPIED, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, and UNITED KINGDOM.
- Lot
- software versions 6.2, 6.2.1, 6.3, and 6.3.1
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity PACS Software Version 7.0 SP0.0.4.7August 3, 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.July 27, 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images andJuly 13, 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 2019July 6, 2022
- FDAE-Z Clean® Disposable Laparoscopic Electrode Model Number 0100L and 0100LS. Expiration dates of 05/2018 toSeptember 4, 2013
- FDAStryker 1) Asnis III Cannulated Screw REF 604640 4.0x40mm TL 13.5mm 2) Asnis III Cannulated Screw REF 604642SSeptember 4, 2013
Questions about this recall
Was GE Ultrasound ViewPoint 6. Product Usage: ViewPoint is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images and generation of medical reports primarily for diagnostic ultrasound. ViewPoint provides the user the ability to including images, drawings, and charts into medical reports. ViewPoint is designed to accept, transfer, display, calculate, store and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint displays to the user, can be used for diagnostic purposes. ViewPoint is intended for professional use only. ViewPoint is not intended to be used as an automated diagnosis system. recalled?
Yes. FDA published recall Z-2098-2013 on September 4, 2013. Reported hazard: GE has recently become aware of a potential safety issue due to the incorrect calculation of the pulsatility index value (PIV) for venous flow measurements associated with obstetric reporting of the GE ViewPoint 6 system. The ViewPoint calculated Doppler PIV for veins could be incorrect. An incorrect calculation of the ductus venosus PIV could lead to false high risk for trisomy 13, 18, and 21,
What should I do if I have this product?
Voluntary: Firm initiated · Letter