Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The device is a component of the MPS Console used as replacements for worn or damaged MPS Console top covers)

Published June 3, 2020 · Recall initiated: December 17, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Quest Medical, Inc.
Distributor
US Nationwide distribution including in the states of NY, PA, TN, WV. OUS: None
Lot
Lot Number: 059135
Risk classification
Class II
Area (FDA)
device
Country
United States