VACUETTE Blood Transfer Unit. Lot no. A19024NX. Expiration date: 2022-02-07 - Product Usage: The Blood Transfer Unit is used to transfer blood from a syringe into the evacuated blood collection tube.
Published June 3, 2020 · Recall initiated: April 3, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
While transferring blood the needle may dislodged and remain in the tube stopper which may cause needle stick injury or blood leakage.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- C.A. Greiner & Sohne Gesellschaftmbh
- Distributor
- *requested 4/15/2020
- Lot
- Item no. 450225. UDI no. 39120017577547 and 29120017577540.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Austria