Big Bore Radiology tomography X-ray system scanner, model number 728244. The Big Bore Radiology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Published July 6, 2016 · Recall initiated: June 14, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom
- Lot
- Big Bore Radiology System Code 728244; Serial numbers 75033, 75110
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States