FDAFDA · Food and drugs·Recall number Z-2110-2016

Big Bore Radiology tomography X-ray system scanner, model number 728244. The Big Bore Radiology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Published July 6, 2016 · Recall initiated: June 14, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom
Lot
Big Bore Radiology System Code 728244; Serial numbers 75033, 75110
Risk classification
Class II
Area (FDA)
device
Country
United States