Big Bore Radiology tomography X-ray system scanner, model number 728244. The Big Bore Radiology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Published July 6, 2016 · Recall initiated: June 14, 2016

Summary

FDA records medical device Class II recall Z-2110-2016, published July 6, 2016 and initiated June 14, 2016. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 89; Lot: Big Bore Radiology System Code 728244; Serial numbers 75033, 75110. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide distribution to Belgium, China, Egypt, Germany, India, Korea, Republic of, Myanmar, Nigeria, Peru, Russian Federation, Rwanda, Singapore, Spain, Sweden, Taiwan, Thailand, and United Kingdom
Lot
Big Bore Radiology System Code 728244; Serial numbers 75033, 75110
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Big Bore Radiology tomography X-ray system scanner, model number 728244. The Big Bore Radiology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. recalled?

Yes. FDA published recall Z-2110-2016 on July 6, 2016. Reported hazard: Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail