Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-011C
Published June 13, 2018 · Recall initiated: April 6, 2018
Summary
FDA records medical device Class II recall Z-2116-2018, published June 13, 2018 and initiated April 6, 2018. Recalling firm: Orthofix, Inc; Status: Terminated; Reported quantity: 50; Lot: UDI: 18056099648592, B1115657, B1107222, B1107223. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Orthofix, Inc
- Distributor
- U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
- Lot
- UDI: 18056099648592, B1115657, B1107222, B1107223
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAY-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b)June 13, 2018
- FDAL-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Left or Right (Part Number): a) LeftJune 13, 2018
- FDABaxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), infusion pump, 35700BAX2 ProductJune 13, 2018
- FDACentricity PACS-IW with Universal Viewer Product Usage: Usage: Centricity PACS-IW with Universal Viewer is aJune 13, 2018
- FDAFirebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A DecorticatingSeptember 20, 2017
- FDAAzure Anterior Cervical Plate System, 2-Level Plate, 36 mm, Implant Grade Titanium and Nitinol, Orthofix Inc.,March 5, 2014
Questions about this recall
Was Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-011C recalled?
Yes. FDA published recall Z-2116-2018 on June 13, 2018. Reported hazard: Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter