Perkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a fully automated, high throughput batch analyzer for time- resolved and prompt fluorescence analysis of samples in microtitration plates. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids. The reagent kits running on the GSP instrument are used to test blood specimens dried on filter paper as an aid in screening newborns for congenital disorders. Currently assays for the determination of T4, TSH, 17OHP, IRT and GALT are cleared and available for use with the GSP instrument.
Published August 8, 2012 · Recall initiated: June 27, 2012
Summary
FDA records medical device Class II recall Z-2121-2012, published August 8, 2012 and initiated June 27, 2012. Recalling firm: Perkinelmer; Status: Terminated; Reported quantity: 78 units; Lot: Serial Numbers: 20210001 through 20210078. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Contamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Perkinelmer
- Distributor
- Worldwide Distribution - US (nationwide) and the countries of Australia, Belgium, Brazil, Egypt, France, Germany, Netherlands, Norway, Saudi Arabia,Switzerland,
- Lot
- Serial Numbers: 20210001 through 20210078
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPerkin Elmer WIZARD2 1-detector,3", 1000 samples Product Code: 2480-0010 The Wizard®" gamma counter isJuly 29, 2015
- FDANEONATAL TOTAL GALACTOSE KIT Product Numbers: 3029-0010, 3029-001B. This kit is intended for the quantitativeJanuary 23, 2013
- FDADBS MICROPLATE, TRUNCATED V-BOTTOMED Product Number: 3033-0010January 23, 2013
- FDANEOGRAM AMINO ACIDS AND ACYLCARNITINES TANDEM MASS SPECTROMETRY KIT Product Number: 3026-0030. This kit isJanuary 23, 2013
- FDAOctane-C Cervical, 6® Lordotic Intended for anterior cervical spine intervertebral body fusion procedures atAugust 8, 2012
- FDAThe Reliance 130 and 130L Cart Washer/Disinfector is a high capacity mechanical washer intended for use in theAugust 8, 2012
Questions about this recall
Was Perkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a fully automated, high throughput batch analyzer for time- resolved and prompt fluorescence analysis of samples in microtitration plates. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids. The reagent kits running on the GSP instrument are used to test blood specimens dried on filter paper as an aid in screening newborns for congenital disorders. Currently assays for the determination of T4, TSH, 17OHP, IRT and GALT are cleared and available for use with the GSP instrument. recalled?
Yes. FDA published recall Z-2121-2012 on August 8, 2012. Reported hazard: Contamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT
What should I do if I have this product?
Voluntary: Firm initiated · Letter