Perkin Elmer Genetic Screening Processor (GSP) Product Code: 2021-0010 Product Usage: The GSP instrument is a fully automated, high throughput batch analyzer for time- resolved and prompt fluorescence analysis of samples in microtitration plates. It is intended for in vitro quantitative / qualitative determination of analytes in body fluids. The reagent kits running on the GSP instrument are used to test blood specimens dried on filter paper as an aid in screening newborns for congenital disorders. Currently assays for the determination of T4, TSH, 17OHP, IRT and GALT are cleared and available for use with the GSP instrument.

Published August 8, 2012 · Recall initiated: June 27, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Contamination from GSP Shaker unit results in decreased signal activity of assay measurements for Genetic Screening for Neonatal IRT, TSH and IRT

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Perkinelmer
Distributor
Worldwide Distribution - US (nationwide) and the countries of Australia, Belgium, Brazil, Egypt, France, Germany, Netherlands, Norway, Saudi Arabia,Switzerland,
Lot
Serial Numbers: 20210001 through 20210078
Risk classification
Class II
Area (FDA)
device
Country
United States