Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
Published May 31, 2017 · Recall initiated: May 15, 2014
Summary
FDA records medical device Class III recall Z-2121-2017, published May 31, 2017 and initiated May 15, 2014. Recalling firm: Medtest Holdings, Inc.; Status: Terminated; Reported quantity: 132919 Vol vials; Lot: Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II). Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtest Holdings, Inc.
- Distributor
- Nationwide
- Lot
- Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II).
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5July 5, 2017
- FDAPointe Scientific Hitachi Total Bilirubin Rl Reagent Product Usage: Two part Chemistry Reagent Hitachi TotalJune 21, 2017
- FDAMedTest DX Assayed Human Multi Sera Control Level 1 (HNC200) and Level 2 (HEC200) HEC200 " Sml amber glassJune 14, 2017
- FDAStatLock® IV Premium Catheter Stabilization Device, product code IV0570. sterile. Packaging: packaged asMay 31, 2017
- FDACSI, Orbital Atherectomy System (OAS) Saline Infusion Pump, Rx only, Part Number: 7-10014, Model Number:May 31, 2017
- FDAPointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PPMay 24, 2017
Questions about this recall
Was Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap. recalled?
Yes. FDA published recall Z-2121-2017 on May 31, 2017. Reported hazard: The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
What should I do if I have this product?
Voluntary: Firm initiated · Letter