Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only
Published July 5, 2017 · Recall initiated: July 21, 2009
Summary
FDA records medical device Class III recall Z-2631-2017, published July 5, 2017 and initiated July 21, 2009. Recalling firm: Medtest Holdings, Inc.; Status: Terminated; Reported quantity: 323 vials; Lot: Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtest Holdings, Inc.
- Distributor
- Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE
- Lot
- Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAda Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist aJuly 5, 2017
- FDATI Transport Monitor/Module Battery Pack Product Usage: intended for monitoring, displaying, reviewing,July 5, 2017
- FDAPointe Scientific Hitachi Total Bilirubin Rl Reagent Product Usage: Two part Chemistry Reagent Hitachi TotalJune 21, 2017
- FDAMedTest DX Assayed Human Multi Sera Control Level 1 (HNC200) and Level 2 (HEC200) HEC200 " Sml amber glassJune 14, 2017
- FDAPointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: AmberMay 31, 2017
- FDAPointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PPMay 24, 2017
Questions about this recall
Was Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only recalled?
Yes. FDA published recall Z-2631-2017 on July 5, 2017. Reported hazard: Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit