FDAFDA · Food and drugs·Recall number Z-2124-2017

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON® Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine

Published May 31, 2017 · Recall initiated: April 24, 2017

Summary

FDA records medical device Class II recall Z-2124-2017, published May 31, 2017 and initiated April 24, 2017. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 3,158 units total (1,957 units in US); Lot: M609458. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Guatemala, India, Italy, Japan, Mayotte, Mexico, Netherlands, Oman, Philippines, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
Lot
M609458
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON® Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine recalled?

Yes. FDA published recall Z-2124-2017 on May 31, 2017. Reported hazard: Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.

What should I do if I have this product?

Voluntary: Firm initiated · Letter