I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.
Published August 8, 2012 · Recall initiated: January 1, 2011
Summary
FDA records medical device Class II recall Z-2133-2012, published August 8, 2012 and initiated January 1, 2011. Recalling firm: Abbott Point Of Care Inc.; Status: Terminated; Reported quantity: 52,750 cartridges; Lot: List number 03P76-25 Lot number S10246A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Point Of Care Inc.
- Distributor
- USA ( nationwide )
- Lot
- List number 03P76-25 Lot number S10246A
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAi-STAT EG6+ cartridge; List Number: 03P77-25;May 6, 2026
- FDAi-STAT G3+ cartridge; List Number: 03P78-26;May 6, 2026
- FDAi-STAT EG7+ cartridge; List Number: 03P76-25;May 6, 2026
- FDAi-STAT EG6+ cartridge. List Number: 03P77-25.September 24, 2025
- FDAExtraneal and/or DIANEAL LOW CALCIUM Peritoneal Dialysis Solution in Plastic Container, PL-146 used withAugust 8, 2012
- FDARF Cannula, 100 mm,Curved 5 MM, 20 G, REF 0406-630-115 RX only, Disposable ( 10/pk) Product Usage: The StrykerAugust 8, 2012
Questions about this recall
Was I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood. recalled?
Yes. FDA published recall Z-2133-2012 on August 8, 2012. Reported hazard: i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
What should I do if I have this product?
Voluntary: Firm initiated · Letter