VITROS 3600 Immunodiagnostic System --- For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents. The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).
Published August 13, 2014 · Recall initiated: July 7, 2014
Summary
FDA records medical device Class II recall Z-2139-2014, published August 13, 2014 and initiated July 7, 2014. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: Domestic: 142, Foreign: 536; Lot: Serial Numbers J36000101 - J36000806 (J Numbers are analogous to Serial Numbers and for VITROS 3600 systems the J Numbers start with 3600XXXX). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
- Lot
- Serial Numbers J36000101 - J36000806 (J Numbers are analogous to Serial Numbers and for VITROS 3600 systems the J Numbers start with 3600XXXX)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDAAdvanDx Quick FISH Fixation Kit for BC ( Blood Cultures), 50 tests. In vitro diagnostic. The QuickFISHAugust 13, 2014
- FDAVITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VITROS 4600 Chemistry System isAugust 13, 2014
Questions about this recall
Was VITROS 3600 Immunodiagnostic System --- For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents. The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID). recalled?
Yes. FDA published recall Z-2139-2014 on August 13, 2014. Reported hazard: Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.
What should I do if I have this product?
Voluntary: Firm initiated · Letter