Back to search
FDAFDA · Food and drugs·Recall number Z-2139-2024

Aeos Robotic Digital Microscope, Product Code: PV010

Published June 19, 2024 · Recall initiated: May 13, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Aesculap Inc
Distributor
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
Lot
Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.