Aeos Robotic Digital Microscope, Product Code: PV010

Published June 19, 2024 · Recall initiated: May 13, 2024

Summary

FDA records medical device Class II recall Z-2139-2024, published June 19, 2024 and initiated May 13, 2024. Recalling firm: Aesculap Inc; Status: Ongoing; Reported quantity: 19 units; Lot: Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aesculap Inc
Distributor
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
Lot
Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Aeos Robotic Digital Microscope, Product Code: PV010 recalled?

Yes. FDA published recall Z-2139-2024 on June 19, 2024. Reported hazard: Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

What should I do if I have this product?

Voluntary: Firm initiated