FDAFDA · Food and drugs·Recall number Z-2141-2024
US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1
Published June 19, 2024 · Recall initiated: May 16, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Distributor
- US Nationwide distribution in the states of AL, IN, AK, WI.
- Lot
- UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087
- Risk classification
- Class II