FDAFDA · Food and drugs·Recall number Z-2141-2024

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Published June 19, 2024 · Recall initiated: May 16, 2024

Summary

FDA records medical device Class II recall Z-2141-2024, published June 19, 2024 and initiated May 16, 2024. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Ongoing; Reported quantity: 4 units; Lot: UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fresenius Medical Care Holdings, Inc.
Distributor
US Nationwide distribution in the states of AL, IN, AK, WI.
Lot
UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1 recalled?

Yes. FDA published recall Z-2141-2024 on June 19, 2024. Reported hazard: Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

What should I do if I have this product?

Voluntary: Firm initiated · Letter