US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1
Published June 19, 2024 · Recall initiated: May 16, 2024
Summary
FDA records medical device Class II recall Z-2141-2024, published June 19, 2024 and initiated May 16, 2024. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Ongoing; Reported quantity: 4 units; Lot: UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Distributor
- US Nationwide distribution in the states of AL, IN, AK, WI.
- Lot
- UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HDJuly 15, 2026
- FDABicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;June 17, 2026
- FDABicarby Dialysate; Model number: RFP-400-G;June 17, 2026
- FDADELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for IntraperitonealMay 13, 2026
- FDAAeos Robotic Digital Microscope, Product Code: PV010June 19, 2024
- FDAVITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 XJune 19, 2024
Questions about this recall
Was US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1 recalled?
Yes. FDA published recall Z-2141-2024 on June 19, 2024. Reported hazard: Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
What should I do if I have this product?
Voluntary: Firm initiated · Letter