INSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream Health Inc.

Published August 13, 2014 · Recall initiated: July 2, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Indicator Dot on the film packet's outer printed paper labeling was incorrectly positioned relative to the location of the raised dot on the dental intraoral film. This may cause confusion for the dental practitioner. If the processed film is placed on the view box by referencing the film's raised dot, the image orientation will be rotated 180 degrees.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Carestream Health Inc.
Distributor
Worldwide Distribution.
Lot
Lot Numbers: 53901201, 53801303, 53801303-1, 54601201, 54601201-1
Risk classification
Class III
Area (FDA)
device
Country
United States