Back to search
FDAFDA · Food and drugs·Recall number Z-2145-2024

VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues. PRODUCT CODE: VCP196H VCP417H VCP602H VCP603H VCP662H VCPB259H VCPB946H

Published June 26, 2024 · Recall initiated: May 10, 2024

Summary

FDA records medical device Class II recall Z-2145-2024, published June 26, 2024 and initiated May 10, 2024. Recalling firm: Ethicon, Inc.; Status: Ongoing; Reported quantity: 55296 units; Lot: GTIN: 10705031051546 10705031052147 10705031053038 10705031054370 10705031054394 10705031054677 10705031150492 10705031056329 10705031056725 10705031150119 10705031150287 10705031150447 PRODUCT CODE PRODUCT LOTS VCP196H UBMDDB VCP417H UAMQML VCP602H UAMQQL UAMSCA VCP603H UAMHTC UAMHZZ UAMEDC VCP662H UBMJSH VCPB259H UAMPKR VCPB946H UAMLJA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ethicon, Inc.
Distributor
Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
Lot
GTIN: 10705031051546 10705031052147 10705031053038 10705031054370 10705031054394 10705031054677 10705031150492 10705031056329 10705031056725 10705031150119 10705031150287 10705031150447 PRODUCT CODE PRODUCT LOTS VCP196H UBMDDB VCP417H UAMQML VCP602H UAMQQL UAMSCA VCP603H UAMHTC UAMHZZ UAMEDC VCP662H UBMJSH VCPB259H UAMPKR VCPB946H UAMLJA
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues. PRODUCT CODE: VCP196H VCP417H VCP602H VCP603H VCP662H VCPB259H VCPB946H recalled?

Yes. FDA published recall Z-2145-2024 on June 26, 2024. Reported hazard: Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

What should I do if I have this product?

Voluntary: Firm initiated · Letter