Human Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The Human Assayed Multi-sera is for the control of accuracy.

Published August 14, 2019 · Recall initiated: June 11, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Randox Laboratories, Limited
Distributor
US Nationwide Distribution - AZ CA CT FL GA IL KY LA MA MD MI MN MO NJ NY OH OK OR PA TX UT VA WA WI WV and PR. Worldwide foreign distribution.
Lot
1306UN, 1308UN, 1309UN
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom