Human Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The Human Assayed Multi-sera is for the control of accuracy.
Published August 14, 2019 · Recall initiated: June 11, 2019
Summary
FDA records medical device Class II recall Z-2150-2019, published August 14, 2019 and initiated June 11, 2019. Recalling firm: Randox Laboratories, Limited; Status: Terminated; Reported quantity: 149 (US); Lot: 1306UN, 1308UN, 1309UN. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Randox Laboratories, Limited
- Distributor
- US Nationwide Distribution - AZ CA CT FL GA IL KY LA MA MD MI MN MO NJ NY OH OK OR PA TX UT VA WA WI WV and PR. Worldwide foreign distribution.
- Lot
- 1306UN, 1308UN, 1309UN
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDARandox Liquid Cardiac Controls Catalogue Numbers CQ5053.November 27, 2019
- FDARANDOX NEFA (Non-Esterified Fatty Acids) FA115 Lot 485343, GTIN: 5055273203066.November 20, 2019
- FDAYXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testingAugust 14, 2019
- FDASenTec Standard Starter Set with and without Service Gas (2 changers each) - Product Usage: Is indicated forAugust 14, 2019
- FDAYXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testingAugust 14, 2019
- FDAMagnesium on RX Imola analyser IVDDecember 14, 2016
Questions about this recall
Was Human Assayed Multi-Sera Level 2, Model NO. HN1530 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The Human Assayed Multi-sera is for the control of accuracy. recalled?
Yes. FDA published recall Z-2150-2019 on August 14, 2019. Reported hazard: The RX Series control target and range value for ALT (Tris buffer without P5P 37C), Bicarbonate and Gamma-GT has been incorrectly assigned in specific lots of the Human Assayed Multi-Sera Controls.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail