HRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900
Published June 26, 2024 · Recall initiated: May 23, 2024
Summary
FDA records medical device Class II recall Z-2152-2024, published June 26, 2024 and initiated May 23, 2024. Recalling firm: Howmedica Osteonics Corp.; Status: Ongoing; Reported quantity: 630 units; Lot: Part 6210-9-900, UDI-DI: 07613327217902, Lot#: All lots back to 2006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Howmedica Osteonics Corp.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, Denmark, Hong Kong, Iran, Italy, Japan, Kuwait, Mexico, New Zealand, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, UK, Venlo.
- Lot
- Part 6210-9-900, UDI-DI: 07613327217902, Lot#: All lots back to 2006.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.October 21, 2026
- FDATriathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.October 21, 2026
- FDATS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.October 21, 2026
- FDATS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.October 21, 2026
- FDACentrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient supportJune 26, 2024
- FDAGray Revision Instrument System Accessory Tray/Case, Catalog/Part Number: 6210-9-910June 26, 2024
Questions about this recall
Was HRIS Storage & Sterilization Case and Tray, Catalog/Part Number: 6210-9-900 recalled?
Yes. FDA published recall Z-2152-2024 on June 26, 2024. Reported hazard: The design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
What should I do if I have this product?
Voluntary: Firm initiated · Letter