FDAFDA · Food and drugs·Recall number Z-2153-2012

Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.

Published August 15, 2012 · Recall initiated: July 2, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Varian Medical Systems, Inc. Oncology Systems
Distributor
Nationwide Distribution including the states of New York and Ohio.
Lot
Serial Numbers/CODES: HIT0535, HIT2953
Risk classification
Class II
Area (FDA)
device
Country
United States