SENSE XL TORSO COIL 1.5T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567141882, 453567141883. Coils may be included in kits with Mode Numbers: 989603014351 & 989603014352.
Published July 3, 2024 · Recall initiated: June 5, 2024
Summary
FDA records medical device Class I recall Z-2163-2024, published July 3, 2024 and initiated June 5, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 729 units; Lot: UDI-DI: N/A. 1. Model Number 453567141882. Serial Numbers: 137, 141, 145, 147, 157, 18, 188, 19, 191, 197, 251, 291, 301, 320, 356, 360, 364, 366, 371, 39, 394, 40, 408, 418, 428, 434, 446, 464, 468, 470, 471, 472, 557, 95, 124, 352, 303, 328, 315, 443, 2, 412, 215, 358, 300, 427, 175, 421, 300202352, 432, 429, 469, 313, 365. 2. Model Number 453567141883. Serial Numbers: 113, 134, 137, 140, 152, 153, 174, 196, 197, 20, 204, 205, 228, 234, 235, 26, 37, 40, 41, 42, 44, 70, 92, 93, 95, 80, 52, 162, 24, 116, 58, 82, 47, 96, 46, 15. 3. Model Number 989603014351. Serial Numbers: IBA_CD_2012, 629, 240, 530, 203, 408, 494, 638. 4. Model Number 989603014352. Serial Numbers: 288, 843, 743, 895, 1169, 1204, 1262, 1003, 500, 1197, 1150, 921, 1063, 1107, 1098, 1120, 1138, 1133, 1139, 1155, 1170, 1156, 1177, 1183, 1187, 1190, 1205, 1212, 1298, 1244, 1247, 1269, 1266, 1274, 1299, 1315, 1283, 457, 844, 1181, 1224, 1258, 1257, 1249, 1073, 440, 465, 429, 1289, 1252, 334, 363, 980, 928, 1092, 1323, 709, 614, 192, 540, 708, 647, 1246, 711, 754, 816, 569, 896, 656, 415, 550, 702, 574, 359, 1147, 712, 1186, 256, 954, 988, 1035, 1055, 411, 968, 327, 586, 819, 450, 1223, 937, 923, 470, 1288, 634, 1027, 715, 1242, 885, 471, 468, 399, 1038, 518, 612, 306, 1117, 661, 289, 1234, 637, 1116, 489, 684, 678, 681, 1128, 813, 1278, 887, 1051, 562, 834, 483, 1113, 773, 1091, RS416, 1185, 662, 568, 572, 1238, 1232, 1094, 180, 1235, 793, 677, 682, 267, 796, 810, 777, 906, 349, 328, 410, 594, 414, 185, 791, 950, 526, 305, 549, 720, 1124, 331, 802, 528, 1200, 663, 343, 706, 753, 458, 539, 650, 600, 1085, 557, 655, 615, 728, 913, 1324, 506, 1049, 1157, 504, 536, 424, 689, 752, 613, 184, 760, 181, 566, 908, 1325, 396, 879, 853, 1194, 1078, 405, 1292, 861, 644, 395, 962, 996, 792, 762, 651, 841, 1062, 246, 302, 840, 982, 1207, 1229, 1074, 1162, 496, 804, 739, 849, 517, 719, 441, 798, 660, 953, 1167, 1225, 1219, 1294, 498, 1182, 1184, 1103, 1202, 674, 693, 960, 1149, 606, 1018, 858, 989, 1259, 878, 1312, 1215, 778, 1037, 1295, 700, 1080, 931, 1281, 278, 1058, 924, 951, 784, 1104, 499, 790, 869, 868, 344, 890, 476, 313, 299, 1189, 1125, 1284, 617, 1168, 666, 1129, 428, 1148, 375, 460, 597, 1302, 1303, 454, 867, 197, 1196, 1280, 1272, 1277, 947, 332, 485, 871, 765, 735, 725, 1217, 1108, 1253, 511, 687, 883, 935, 1311, 909, 899, 444, 377, 824, 341, 742, 654, 856, 290, 1203, 701, 970, 1060, 710, 473, 1064, 218, 1008, 894, 1045, 975, 1054, 665, 806, 783, 704, 547, 1301, 1135, 800, 939, 695, 199, 205, 1241, 963, 209, 1059, 916, 605, 204, 308, 749, 516, 1119, 337, 748, 1095, 717, 925, 832, 966, 683, 734, 977, 942, 1046, 587, 671, 339, 1021, 385, 1270, 995, 976, 653, 623, 884, 1132, 741, 1195, 945, 1171, 865, 215, 774, 991, 955, 1114, 325, 262, 434, 1254, 1015, 372, 281, 635, 667, 1005, 639, 585, 1122, 537, 676, 439, 543, 488, 697, 383, 964, 1172, 904, 857, 1016, 559, 565, 736, 626, 625, 679, 624, 973, 675, 696, 652, 721, 690, 862, 874, 915, 911, 771, 912, 1158, 999, 892, 1028, 756, 948, 632, 998, 1211, 1048, 1255, 781, 738, 932, 1033, 1042, 936, 1075, 917, 1123, 919, 855, 956, 927, 1050, 889, 910, 891, 1209, 946, 852, 1066, 901, 1230, 967, 1118, 459, 1034, 1099, 1070, 990, 1077, 1072, 978, 961, 801, 1039, 1071, 1061, 1263, 983, 877, 926, 1069, 974, 1032, 1040, 1052, 342, 872, 1086, 1020, 1228, 888, 822, 1250, 1100, 1068, 1293, 831, 1314, 1233, 1240, 1251, 336, 992, 713, 1227, 729, 882, 1319, 406, 1067, 1213, 1286, 922, 993, 522, 758, 714, 1214, 859, 987, 694, 907, 668, 789, 1141, 902, 1102, 464, 1271, 1081, 794, 645, 699, 576, 972, 1276, 376, 315, 827, 1089, 905, 836, 997, 1275, 1206, 698, 830, 930, 1083, 479, 1121, 669, 1140, 1191, 257, 914, 558, 751, 1096, 1090, 805, 1267, 692, 940, 823, 1024, 1101, 1014, 846, 1176, 553, 618, 409, 1126, 447, 456, 1151, 231, 680, 971, 965, 817, 222, 1239, 1265, 691, 1164, 938, 803, 412. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for coils to heat up and harm patients (burn).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Papua New Guinea, Peru, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, T®rkiye, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
- Lot
- UDI-DI: N/A. 1. Model Number 453567141882. Serial Numbers: 137, 141, 145, 147, 157, 18, 188, 19, 191, 197, 251, 291, 301, 320, 356, 360, 364, 366, 371, 39, 394, 40, 408, 418, 428, 434, 446, 464, 468, 470, 471, 472, 557, 95, 124, 352, 303, 328, 315, 443, 2, 412, 215, 358, 300, 427, 175, 421, 300202352, 432, 429, 469, 313, 365. 2. Model Number 453567141883. Serial Numbers: 113, 134, 137, 140, 152, 153, 174, 196, 197, 20, 204, 205, 228, 234, 235, 26, 37, 40, 41, 42, 44, 70, 92, 93, 95, 80, 52, 162, 24, 116, 58, 82, 47, 96, 46, 15. 3. Model Number 989603014351. Serial Numbers: IBA_CD_2012, 629, 240, 530, 203, 408, 494, 638. 4. Model Number 989603014352. Serial Numbers: 288, 843, 743, 895, 1169, 1204, 1262, 1003, 500, 1197, 1150, 921, 1063, 1107, 1098, 1120, 1138, 1133, 1139, 1155, 1170, 1156, 1177, 1183, 1187, 1190, 1205, 1212, 1298, 1244, 1247, 1269, 1266, 1274, 1299, 1315, 1283, 457, 844, 1181, 1224, 1258, 1257, 1249, 1073, 440, 465, 429, 1289, 1252, 334, 363, 980, 928, 1092, 1323, 709, 614, 192, 540, 708, 647, 1246, 711, 754, 816, 569, 896, 656, 415, 550, 702, 574, 359, 1147, 712, 1186, 256, 954, 988, 1035, 1055, 411, 968, 327, 586, 819, 450, 1223, 937, 923, 470, 1288, 634, 1027, 715, 1242, 885, 471, 468, 399, 1038, 518, 612, 306, 1117, 661, 289, 1234, 637, 1116, 489, 684, 678, 681, 1128, 813, 1278, 887, 1051, 562, 834, 483, 1113, 773, 1091, RS416, 1185, 662, 568, 572, 1238, 1232, 1094, 180, 1235, 793, 677, 682, 267, 796, 810, 777, 906, 349, 328, 410, 594, 414, 185, 791, 950, 526, 305, 549, 720, 1124, 331, 802, 528, 1200, 663, 343, 706, 753, 458, 539, 650, 600, 1085, 557, 655, 615, 728, 913, 1324, 506, 1049, 1157, 504, 536, 424, 689, 752, 613, 184, 760, 181, 566, 908, 1325, 396, 879, 853, 1194, 1078, 405, 1292, 861, 644, 395, 962, 996, 792, 762, 651, 841, 1062, 246, 302, 840, 982, 1207, 1229, 1074, 1162, 496, 804, 739, 849, 517, 719, 441, 798, 660, 953, 1167, 1225, 1219, 1294, 498, 1182, 1184, 1103, 1202, 674, 693, 960, 1149, 606, 1018, 858, 989, 1259, 878, 1312, 1215, 778, 1037, 1295, 700, 1080, 931, 1281, 278, 1058, 924, 951, 784, 1104, 499, 790, 869, 868, 344, 890, 476, 313, 299, 1189, 1125, 1284, 617, 1168, 666, 1129, 428, 1148, 375, 460, 597, 1302, 1303, 454, 867, 197, 1196, 1280, 1272, 1277, 947, 332, 485, 871, 765, 735, 725, 1217, 1108, 1253, 511, 687, 883, 935, 1311, 909, 899, 444, 377, 824, 341, 742, 654, 856, 290, 1203, 701, 970, 1060, 710, 473, 1064, 218, 1008, 894, 1045, 975, 1054, 665, 806, 783, 704, 547, 1301, 1135, 800, 939, 695, 199, 205, 1241, 963, 209, 1059, 916, 605, 204, 308, 749, 516, 1119, 337, 748, 1095, 717, 925, 832, 966, 683, 734, 977, 942, 1046, 587, 671, 339, 1021, 385, 1270, 995, 976, 653, 623, 884, 1132, 741, 1195, 945, 1171, 865, 215, 774, 991, 955, 1114, 325, 262, 434, 1254, 1015, 372, 281, 635, 667, 1005, 639, 585, 1122, 537, 676, 439, 543, 488, 697, 383, 964, 1172, 904, 857, 1016, 559, 565, 736, 626, 625, 679, 624, 973, 675, 696, 652, 721, 690, 862, 874, 915, 911, 771, 912, 1158, 999, 892, 1028, 756, 948, 632, 998, 1211, 1048, 1255, 781, 738, 932, 1033, 1042, 936, 1075, 917, 1123, 919, 855, 956, 927, 1050, 889, 910, 891, 1209, 946, 852, 1066, 901, 1230, 967, 1118, 459, 1034, 1099, 1070, 990, 1077, 1072, 978, 961, 801, 1039, 1071, 1061, 1263, 983, 877, 926, 1069, 974, 1032, 1040, 1052, 342, 872, 1086, 1020, 1228, 888, 822, 1250, 1100, 1068, 1293, 831, 1314, 1233, 1240, 1251, 336, 992, 713, 1227, 729, 882, 1319, 406, 1067, 1213, 1286, 922, 993, 522, 758, 714, 1214, 859, 987, 694, 907, 668, 789, 1141, 902, 1102, 464, 1271, 1081, 794, 645, 699, 576, 972, 1276, 376, 315, 827, 1089, 905, 836, 997, 1275, 1206, 698, 830, 930, 1083, 479, 1121, 669, 1140, 1191, 257, 914, 558, 751, 1096, 1090, 805, 1267, 692, 940, 823, 1024, 1101, 1014, 846, 1176, 553, 618, 409, 1126, 447, 456, 1151, 231, 680, 971, 965, 817, 222, 1239, 1265, 691, 1164, 938, 803, 412.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDAInspire Model 3028, IV Implantable Pulse GeneratorJuly 3, 2024
- FDASENSE XL TORSO COIL 3.0T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera andJuly 3, 2024
Questions about this recall
Was SENSE XL TORSO COIL 1.5T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567141882, 453567141883. Coils may be included in kits with Mode Numbers: 989603014351 & 989603014352. recalled?
Yes. FDA published recall Z-2163-2024 on July 3, 2024. Reported hazard: Potential for coils to heat up and harm patients (burn).
What should I do if I have this product?
Voluntary: Firm initiated · Letter