SENSE XL TORSO COIL 3.0T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567394941, 453567394942, 453567394943, 453567394945, 459801593182. Coils may be included in kits with Mode Numbers: 989603050641, 989603050642.

Published July 3, 2024 · Recall initiated: June 5, 2024

Summary

FDA records medical device Class I recall Z-2164-2024, published July 3, 2024 and initiated June 5, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 307 units; Lot: UDI-DI: N/A. 1. Model Number 453567394941. Serial Number: 9. 2. Model Number 453567394942. Serial Numbers: 40000, 100, 102, 104, 144, 192, 84, 98, 42. 3. Model Number 453567394943. Serial Numbers: 301, 228, 761, 87, 880, 814, 0174R3, 541, @001@0179R3, 0141R3, 0172R3, 0189R3, 0260R3, 0265R3, 0285R3, 0358R3, 116R3, 121, 127, 149, 152, 158, 165, 176, 180, 213R3, 216, 220, 224, 235, 237, 263, 265, 267, 273, 274, 286, 294, 298, 302, 303, 310R3, 312, 313, 316, 328R3, 336, 346, 359, 360, 374, 377, 382, 389, 400, 406, 421, 423, 424, 43, 431, 45, 454, 455, 460, 464, 466, 474, 477, 481, 488, 496, 505, 511, 513, 518, 523, 543, 549, 564, 576, 582, 592, 597, 602, 603, 61, 610, 611, 617R3, 627, 632, 633, 635, 642, 657, 658, 666r3, 669, 670, 672, 685, 686, 694, 708, 710, 716, 720, 726, 729, 732, 736, 745, 757, 761, 763, 780, 782, 783, 788, 793, 800, 828, 84, 861, 87, 880, 887, 89, 907, 908, 913, 924, 926, 932, 939, 947, 956, 961, 967. 4. Model Number 453567394945. Serial Numbers: 104, 107, 108, 117, 129, 145, 165, 166, 178, 180, 198, 201, 211, 22, 222, 225, 229, 242, 268, 31, 336, 35, 42, 64, 69, 87, 916, 94. 5. Model Number 459801593182. Serial Numbers: 116, 141, 165, 211, 100, 103, 104, 110, 112, 123, 124, 127, 128, 13, 131, 133, 134, 136, 142, 143, 144, 145, 146, 148, 15, 150, 156, 158, 164, 167, 168, 172, 173, 174, 179, 183, 186, 192, 193, 194, 195, 198, 199, 20, 204, 208, 21, 217, 221, 225, 23, 230, 232, 240, 241, 246, 249, 251, 253, 254, 26, 262, 264, 265, 266, 268, 275, 278, 279, 28, 288, 294, 297, 298, 303, 306, 311, 312, 313, 315, 321, 324, 326, 330, 339, 34, 340, 344, 345, 346, 349, 350, 352, 354, 355, 357, 361, 362, 365, 368, 370, 39, 40, 42586, 44, 52, 53, 57, 6, 60, 62, 63, 64, 66, 705, 71, 72, 79, 80, 88, 932, 94, 97. 6. Model Number 989603050641. Serial Numbers: 152, 169, 179, 88, 83, 109, 105, 130, 176. 7. Model Number 989603050642. Serial Numbers: 529, 509, 217, 189, 127, 311, 85, 531, 534, 546, 561, 347, 609, 296, 142, 110, 391, 137, 20, 28, 160, 378, 370, 172, 202, 76, 272, 326, 404, 167, 94, 446, 560, 521, 497, 353, 515, 386, 369, 253, 571, 220, 333, 362, 265, 426, 442, 469, 74, 31, 284, 533, 147, 248, 259, 69, 330, 270, 525, 611, 102, 309, 461, 52, 191, 203, 108, 8, 100, 524, 381, 503, 389, 283, 360, 53, 348, 448, 490, 556, 22, 380, 247, 59, 554, 589, 580, 403, 507, 226, 452, 183, 357, 224, 472, 39, 221, 435, 16, 480, 266, 255, 445, 588, 1, 128, 466, 32, 124, 133, 276, 329, 431, 443, 598, 227, 584, 14, 449, 295, 513, 168, 535, 40, 17, 48, 104, 51, 177, 64, 287, 9, 394, 174, 494, 201, 50, 482, 590, 350, 438, 487, 365, 297, 368, 57, 196, 432, 551, 310, 12, 242, 250, 249, 240, 366, 382, 18, 390, 320, 129, 604, 314, 545, 437, 58, 583, 175, 367, 274, 510, 512, 573, 131, 505, 44, 599, 47, 595, 398, 206, 593, 564, 495, 619, 54, 30, 122, 552, 312, 161, 254, 275, 282, 387, 123, 335, 364, 570, 607, 166, 91, 384, 379, 436, 508, 193, 474, 407, 575, 577, 608, 601, 356, 188, 232, 324, 430, 422, 455, 155, 195, 321, 300, 361, 194, 208, 338, 418, 214, 99, 243, 492, 317, 581, 103, 237, 340, 401. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for coils to heat up and harm patients (burn).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution: US (Nationwide) and OUS (Foreign): Afghanistan, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Papua New Guinea, Peru, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, T®rkiye, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
Lot
UDI-DI: N/A. 1. Model Number 453567394941. Serial Number: 9. 2. Model Number 453567394942. Serial Numbers: 40000, 100, 102, 104, 144, 192, 84, 98, 42. 3. Model Number 453567394943. Serial Numbers: 301, 228, 761, 87, 880, 814, 0174R3, 541, @001@0179R3, 0141R3, 0172R3, 0189R3, 0260R3, 0265R3, 0285R3, 0358R3, 116R3, 121, 127, 149, 152, 158, 165, 176, 180, 213R3, 216, 220, 224, 235, 237, 263, 265, 267, 273, 274, 286, 294, 298, 302, 303, 310R3, 312, 313, 316, 328R3, 336, 346, 359, 360, 374, 377, 382, 389, 400, 406, 421, 423, 424, 43, 431, 45, 454, 455, 460, 464, 466, 474, 477, 481, 488, 496, 505, 511, 513, 518, 523, 543, 549, 564, 576, 582, 592, 597, 602, 603, 61, 610, 611, 617R3, 627, 632, 633, 635, 642, 657, 658, 666r3, 669, 670, 672, 685, 686, 694, 708, 710, 716, 720, 726, 729, 732, 736, 745, 757, 761, 763, 780, 782, 783, 788, 793, 800, 828, 84, 861, 87, 880, 887, 89, 907, 908, 913, 924, 926, 932, 939, 947, 956, 961, 967. 4. Model Number 453567394945. Serial Numbers: 104, 107, 108, 117, 129, 145, 165, 166, 178, 180, 198, 201, 211, 22, 222, 225, 229, 242, 268, 31, 336, 35, 42, 64, 69, 87, 916, 94. 5. Model Number 459801593182. Serial Numbers: 116, 141, 165, 211, 100, 103, 104, 110, 112, 123, 124, 127, 128, 13, 131, 133, 134, 136, 142, 143, 144, 145, 146, 148, 15, 150, 156, 158, 164, 167, 168, 172, 173, 174, 179, 183, 186, 192, 193, 194, 195, 198, 199, 20, 204, 208, 21, 217, 221, 225, 23, 230, 232, 240, 241, 246, 249, 251, 253, 254, 26, 262, 264, 265, 266, 268, 275, 278, 279, 28, 288, 294, 297, 298, 303, 306, 311, 312, 313, 315, 321, 324, 326, 330, 339, 34, 340, 344, 345, 346, 349, 350, 352, 354, 355, 357, 361, 362, 365, 368, 370, 39, 40, 42586, 44, 52, 53, 57, 6, 60, 62, 63, 64, 66, 705, 71, 72, 79, 80, 88, 932, 94, 97. 6. Model Number 989603050641. Serial Numbers: 152, 169, 179, 88, 83, 109, 105, 130, 176. 7. Model Number 989603050642. Serial Numbers: 529, 509, 217, 189, 127, 311, 85, 531, 534, 546, 561, 347, 609, 296, 142, 110, 391, 137, 20, 28, 160, 378, 370, 172, 202, 76, 272, 326, 404, 167, 94, 446, 560, 521, 497, 353, 515, 386, 369, 253, 571, 220, 333, 362, 265, 426, 442, 469, 74, 31, 284, 533, 147, 248, 259, 69, 330, 270, 525, 611, 102, 309, 461, 52, 191, 203, 108, 8, 100, 524, 381, 503, 389, 283, 360, 53, 348, 448, 490, 556, 22, 380, 247, 59, 554, 589, 580, 403, 507, 226, 452, 183, 357, 224, 472, 39, 221, 435, 16, 480, 266, 255, 445, 588, 1, 128, 466, 32, 124, 133, 276, 329, 431, 443, 598, 227, 584, 14, 449, 295, 513, 168, 535, 40, 17, 48, 104, 51, 177, 64, 287, 9, 394, 174, 494, 201, 50, 482, 590, 350, 438, 487, 365, 297, 368, 57, 196, 432, 551, 310, 12, 242, 250, 249, 240, 366, 382, 18, 390, 320, 129, 604, 314, 545, 437, 58, 583, 175, 367, 274, 510, 512, 573, 131, 505, 44, 599, 47, 595, 398, 206, 593, 564, 495, 619, 54, 30, 122, 552, 312, 161, 254, 275, 282, 387, 123, 335, 364, 570, 607, 166, 91, 384, 379, 436, 508, 193, 474, 407, 575, 577, 608, 601, 356, 188, 232, 324, 430, 422, 455, 155, 195, 321, 300, 361, 194, 208, 338, 418, 214, 99, 243, 492, 317, 581, 103, 237, 340, 401.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SENSE XL TORSO COIL 3.0T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567394941, 453567394942, 453567394943, 453567394945, 459801593182. Coils may be included in kits with Mode Numbers: 989603050641, 989603050642. recalled?

Yes. FDA published recall Z-2164-2024 on July 3, 2024. Reported hazard: Potential for coils to heat up and harm patients (burn).

What should I do if I have this product?

Voluntary: Firm initiated · Letter