OPERON D850 Surgical Table Product Usage: Surgical Table
Published September 18, 2013 · Recall initiated: April 16, 2012
Summary
FDA records medical device Class II recall Z-2168-2013, published September 18, 2013 and initiated April 16, 2012. Recalling firm: Berchtold Corp.; Status: Terminated; Reported quantity: 263; Lot: 3501110-S10023, 3501110-S10024, 3501110-S10025, 3501110-S10205, 3501110-S10208, 3501110-S10064, 3501110-S10036, 3501110-S10043, 3501110-S10045, 3501110-S10050, 3501110-S10061, 3501110-S10062, 3511110-S10139, 3511110-S10143, 3511110-S10144, 3511110-S10162, 3501110-S10192, 3501110-S10018, 3511110-S10038, 3511110-S10041, 3511110-S10047, 3511110-S10072, 3511110-S10137, 3511110-S10138, 3501110-S10153, 3501110-S10166, 3501110-S10093, 3501110-S10104, 3501110-S10090, 3501110-S10147, 3501110-S10262, 3501110-S10281, 3501110-S10105, 3501110-S10172, 3501110-S10195, 3501110-S10109, 3501110-S10134, 3501110-S10099, 3501110-S10101, 3501110-S10151, 3511110-S10086, 3511110-S10092, 3511110-S10085, 3511110-S10091, 3501110-S10065, 3501110-S10037, 3501110-S10056, 3501110-S10259, 3501110-S10083, 3501110-S10170, 3501110-S10212, 3501110-S10067, 3501110-S10073, 3501110-S10116, 3501110-S10117, 3501110-S10119, 3501110-S10157, 3501110-S10167, 3501110-S10204, 3501110-S10207, 3501110S-10110, 3511110-S10234, 3501110-S10078, 3501110-S10034, 3501110-S10035, 3501110-S10097, 3501110-S10112, 3501110-S10088, 3501110-S10096, 3501110-S10098, 3501110-S10102, 3501110-S10111, 3501110-S10019, 3501110-S10021, 3501110-S10022, 3501110-S10223, 3501110-S10267, 3501110-S10049, 3501110-S10044, 3501110-S10052, 3501110-S10149, 3501110-S10121, 3501110-S10129, 3501110-S10254, 3501110-S10257, 3501110-S10263, 3501110-S10100, 3511110-S10176, 3511110-S10221, 3501110-S10026, 3501110-S10051, 3501110-S10211, 3501110-S10251, 3501110-S10046, 3501110-S10053, 3501110-S10054, 3501110-S10055, 3501110-S10063, 3501110-S10080, 3501110-S10028, 3501110-S10027, 3501110-S10027, 3501110-S10152, 3511110-S10161, 3501110-S10079, 3501110-S10095, 3511110-S10039, 3501110-S10103, 3501110-S10209, 3501110-S10213, 3501110-S10076, 3501110-S10114, 350110-S10118, 3501110-S10126, 3501110-S10128, 3501110-S10136, 3501110-S10154, 3501110-S10156, 3501110-S10158, 3501110-S10258, 3501110-S10249, 3501110-S10256, 3501110-S10260, 3501110-S10261, 3501110-S10031, 3501110-S10210, 3501110-S10277, 3501110-S10070, 3501110-10082, 35011100-S10084, 3501110-S10087, 3501110-S10089, 3501110-S10248, 3501110-S10250, 3501110-S10131, 3501110-S10094, 3501110-S10016, 3501110-S10017, 3501110-S10020, 3501110-S10030, 3501110-S10206, 3501110-S10252, 3501110-S10148, 3501110-S10155, 3501110-S10171, 3501110-S10173, 3511110-S10107, 3501110-S10193, 35011110-S10233, 3511110-S10159, 3511110-S10160, 3511110-S10168, 3511110-S10169, 3511110-S10181, 3511110-S10196, 3511110-S10197, 3511110-S10198, 3511110-S10199, 3511110-S10200, 3511110-S10201, 3511110-S10202, 3511110-S10214, 3511110-S10215, 3511110-S10218, 3511110-S10219, 3511110-S10228, 3511110-S10235, 3501110-S10253, 3501110-S10255, 3511110-S10033, 3501110-S10074, 3501110-S10075, 3501110-S10077, 3511110-S10040, 3511110-S10145, 3511110-S10032, 3501110-S10270, 3501110-S10135, 3501110-S10150, 3501110-S10164, 3501110-S10165, 3501110-S10265, 3501110-S10266, 3501110-S10280, 3501110-S10060, 3501110-S10133, 3501110-S10106, 3511110-S10108, 3511110-S10217, 3511110-S10220, 3501110-S10113, 3501110-S10115, 3501110-S10120, 3501110-S10122, 3501110-S10123, 3501110-S10125, 3501110-S10130, 3501110-S10132, 3511110-S10140, 3511110-S10141, 3511110-S10142, 3511110-S10146, 3501110-S10066, 3501110-S10068, 3501110-S10069, 3501110-S10081, 3501110-S10222, 3501110-S10224, 3501110-S10225, 3501110-S10226, 3501110-S10227, 3501110-S10230, 3501110-S10231, 3501110-S10236, 3501110-S10237, 3501110-S10238, 3501110-S10239, 3501110-S10071, 3501110-S10124, 3501110-S10127, 3511110-S10177, 3511110-S10178, 3511110-S10179, 3511110-S10180, 3511110-S10203. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Table was not lowering.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Berchtold Corp.
- Distributor
- USA Nationwide Distribution in the states of AL, AZ, CA, CO, CT, FL, GA, HI.IL, IA, KY, LA, MD, MA, MI, MN, MS, NH, NJ, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WI and including Canada.
- Lot
- 3501110-S10023, 3501110-S10024, 3501110-S10025, 3501110-S10205, 3501110-S10208, 3501110-S10064, 3501110-S10036, 3501110-S10043, 3501110-S10045, 3501110-S10050, 3501110-S10061, 3501110-S10062, 3511110-S10139, 3511110-S10143, 3511110-S10144, 3511110-S10162, 3501110-S10192, 3501110-S10018, 3511110-S10038, 3511110-S10041, 3511110-S10047, 3511110-S10072, 3511110-S10137, 3511110-S10138, 3501110-S10153, 3501110-S10166, 3501110-S10093, 3501110-S10104, 3501110-S10090, 3501110-S10147, 3501110-S10262, 3501110-S10281, 3501110-S10105, 3501110-S10172, 3501110-S10195, 3501110-S10109, 3501110-S10134, 3501110-S10099, 3501110-S10101, 3501110-S10151, 3511110-S10086, 3511110-S10092, 3511110-S10085, 3511110-S10091, 3501110-S10065, 3501110-S10037, 3501110-S10056, 3501110-S10259, 3501110-S10083, 3501110-S10170, 3501110-S10212, 3501110-S10067, 3501110-S10073, 3501110-S10116, 3501110-S10117, 3501110-S10119, 3501110-S10157, 3501110-S10167, 3501110-S10204, 3501110-S10207, 3501110S-10110, 3511110-S10234, 3501110-S10078, 3501110-S10034, 3501110-S10035, 3501110-S10097, 3501110-S10112, 3501110-S10088, 3501110-S10096, 3501110-S10098, 3501110-S10102, 3501110-S10111, 3501110-S10019, 3501110-S10021, 3501110-S10022, 3501110-S10223, 3501110-S10267, 3501110-S10049, 3501110-S10044, 3501110-S10052, 3501110-S10149, 3501110-S10121, 3501110-S10129, 3501110-S10254, 3501110-S10257, 3501110-S10263, 3501110-S10100, 3511110-S10176, 3511110-S10221, 3501110-S10026, 3501110-S10051, 3501110-S10211, 3501110-S10251, 3501110-S10046, 3501110-S10053, 3501110-S10054, 3501110-S10055, 3501110-S10063, 3501110-S10080, 3501110-S10028, 3501110-S10027, 3501110-S10027, 3501110-S10152, 3511110-S10161, 3501110-S10079, 3501110-S10095, 3511110-S10039, 3501110-S10103, 3501110-S10209, 3501110-S10213, 3501110-S10076, 3501110-S10114, 350110-S10118, 3501110-S10126, 3501110-S10128, 3501110-S10136, 3501110-S10154, 3501110-S10156, 3501110-S10158, 3501110-S10258, 3501110-S10249, 3501110-S10256, 3501110-S10260, 3501110-S10261, 3501110-S10031, 3501110-S10210, 3501110-S10277, 3501110-S10070, 3501110-10082, 35011100-S10084, 3501110-S10087, 3501110-S10089, 3501110-S10248, 3501110-S10250, 3501110-S10131, 3501110-S10094, 3501110-S10016, 3501110-S10017, 3501110-S10020, 3501110-S10030, 3501110-S10206, 3501110-S10252, 3501110-S10148, 3501110-S10155, 3501110-S10171, 3501110-S10173, 3511110-S10107, 3501110-S10193, 35011110-S10233, 3511110-S10159, 3511110-S10160, 3511110-S10168, 3511110-S10169, 3511110-S10181, 3511110-S10196, 3511110-S10197, 3511110-S10198, 3511110-S10199, 3511110-S10200, 3511110-S10201, 3511110-S10202, 3511110-S10214, 3511110-S10215, 3511110-S10218, 3511110-S10219, 3511110-S10228, 3511110-S10235, 3501110-S10253, 3501110-S10255, 3511110-S10033, 3501110-S10074, 3501110-S10075, 3501110-S10077, 3511110-S10040, 3511110-S10145, 3511110-S10032, 3501110-S10270, 3501110-S10135, 3501110-S10150, 3501110-S10164, 3501110-S10165, 3501110-S10265, 3501110-S10266, 3501110-S10280, 3501110-S10060, 3501110-S10133, 3501110-S10106, 3511110-S10108, 3511110-S10217, 3511110-S10220, 3501110-S10113, 3501110-S10115, 3501110-S10120, 3501110-S10122, 3501110-S10123, 3501110-S10125, 3501110-S10130, 3501110-S10132, 3511110-S10140, 3511110-S10141, 3511110-S10142, 3511110-S10146, 3501110-S10066, 3501110-S10068, 3501110-S10069, 3501110-S10081, 3501110-S10222, 3501110-S10224, 3501110-S10225, 3501110-S10226, 3501110-S10227, 3501110-S10230, 3501110-S10231, 3501110-S10236, 3501110-S10237, 3501110-S10238, 3501110-S10239, 3501110-S10071, 3501110-S10124, 3501110-S10127, 3511110-S10177, 3511110-S10178, 3511110-S10179, 3511110-S10180, 3511110-S10203
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAXlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product isSeptember 18, 2013
- FDAADVIA Centaur Folate (100 Tests), SMN 10310308, Cat No. 06367974 In vitro diagnostic use in the quantitativeSeptember 18, 2013
- FDAOutlook Safety Infusion System Burette Set. For intravenous fluid administration.September 18, 2013
- FDAADVIA Centaur Folate (500 Tests) SMN 10325366, Cat No. 06891541 In vitro diagnostic use in the quantitativeSeptember 18, 2013
- FDAHand Controls for Operon D750, D752, D760, D820 and D850 The hand controls are used to operate OPERON surgicalJune 12, 2013
- FDAChromovision High Definition (HD) Wireless Camera, Pt # CC-9715900. Visual recording of surgical procedures.August 15, 2012
Questions about this recall
Was OPERON D850 Surgical Table Product Usage: Surgical Table recalled?
Yes. FDA published recall Z-2168-2013 on September 18, 2013. Reported hazard: Table was not lowering.
What should I do if I have this product?
Voluntary: Firm initiated · Letter