Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient.
Published July 29, 2015 · Recall initiated: May 28, 2015
Summary
FDA records medical device Class II recall Z-2170-2015, published July 29, 2015 and initiated May 28, 2015. Recalling firm: American Medical Systems Innovation Center - Silicon Valley; Status: Terminated; Reported quantity: 16,710 total all affected devices; Lot: Model 0010-0760; All product manufactured from April 8, 2012 to April 8, 2015. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Medical Systems Innovation Center - Silicon Valley
- Distributor
- Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
- Lot
- Model 0010-0760; All product manufactured from April 8, 2012 to April 8, 2015
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATool, Stripper W/BLD, 18S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protectiveJuly 29, 2015
- FDATool, Stripper W/BLD, 44S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protectiveJuly 29, 2015
- FDATHOMAS by Gardner Denver, MEDI-PUMP(R) Aspirator, Model 1243 Product Usage: medical aspiratorJuly 29, 2015
- FDAVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR ScoutJuly 29, 2015
- FDAPerkin Elmer WIZARD2 1-detector,3", 1000 samples Product Code: 2480-0010 The Wizard®" gamma counter isJuly 29, 2015
- FDAARCHITECT c8000 System List Number 01G06-98 Product Usage: The Abbott ARCHITECT cSystem is intended for InJuly 29, 2015
Questions about this recall
Was Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient. recalled?
Yes. FDA published recall Z-2170-2015 on July 29, 2015. Reported hazard: validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
What should I do if I have this product?
Voluntary: Firm initiated · Letter