Tool, Stripper W/BLD, 44S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.
Published July 29, 2015 · Recall initiated: May 28, 2015
Summary
FDA records medical device Class II recall Z-2181-2015, published July 29, 2015 and initiated May 28, 2015. Recalling firm: American Medical Systems Innovation Center - Silicon Valley; Status: Terminated; Reported quantity: 16,710 total affected devices; Lot: Model 80-10008-002; All product manufactured from April 8, 2012 to April 8, 2015. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Medical Systems Innovation Center - Silicon Valley
- Distributor
- Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
- Lot
- Model 80-10008-002; All product manufactured from April 8, 2012 to April 8, 2015
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAClonidine Hydrochloride Tablets, USP, 0.1 mg, packaged in a) 100 count (NDC 53489-215-01) and b) 1000 countAugust 5, 2015
- FDAClonidine Hydrochloride Tablets, USP, 0.3 mg, 100 count HDPE bottles, Rx only, Mutual Pharmaceutical,August 5, 2015
- FDATool, Stripper W/BLD, 18S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protectiveJuly 29, 2015
- FDAFiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The FiberJuly 29, 2015
- FDAARCHITECT c8000 System List Number 01G06-98 Product Usage: The Abbott ARCHITECT cSystem is intended for InJuly 29, 2015
- FDATHOMAS by Gardner Denver, MEDI-PUMP(R) Aspirator, Model 1243 Product Usage: medical aspiratorJuly 29, 2015
Questions about this recall
Was Tool, Stripper W/BLD, 44S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size. recalled?
Yes. FDA published recall Z-2181-2015 on July 29, 2015. Reported hazard: validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
What should I do if I have this product?
Voluntary: Firm initiated · Letter