EndoWrist Stapler System User Manual Addendum and Wall Chart; The EndoWrist Stapler 45 System and EndoWrist Stapler 45 Reloads are intended to be used with the da Vinci Si Surgical System (Model IS3000) for resection, transection and/or creation of anastomoses in General, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

Published August 20, 2014 · Recall initiated: June 2, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Changes were made to the Stapler EndoWrist Stapler 45 System Instruments and Accessories User Manual Addendum (550986-03) and the EndoWrist Stapler 45 Wall Chart (PN 551524-01) in which the pH range specification for the enzymatic cleaner used throughout the cleaning process was modified from pH-neutral to pH-neutral to mild alkaline (mild alkaline is defined as pHd11). Mild Alkaline enzymatic

What to do

Voluntary: Firm initiated · Visit

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
Distribution to Florida and South Carolina
Lot
Model number 410298-05 and 410298-06
Risk classification
Class II
Area (FDA)
device
Country
United States